9 recommendations / guidelines – KERN & SOHN MLS 50-3C User Manual
Page 76

76
MLS_C-BA-e-1330
11.9 Recommendations / Guidelines
Prepare standard sample:
-
Crush sample, as required, and spread it evenly in the aluminium dish.
Prepare special samples:
-
For sensitive or hard to spread test materials (e. g. mercury) a glass fibre filter
is available for use.
-
Spread sample evenly on glass fibre filter and cover is with a second glass
fibre filter.
-
The glass fibre filter is also useful as a protection when splashing materials
are dealt with (each splash falsifies the final result).
MATERIAL Specimen
weight
(g)
Drying
temperature
(º C)
Data query
interval
(s)
% moisture
% solid
Drying time
(min)
Dried piece
of apple
5-8
100
10
76.5
10-15
Moist apple
5-8
100
10
7.5
5-10
Butter
2-5
138
15
16.3
4.5
Mustard
2-3
130
20
76.4
10
Ground
coffee
2-3
106
5
2.8
4
Cornflakes
2-4
120
15
9.7
5-7
Yoghurt
2-3
110
15
86.5
4.5-6.5
Cocoa
powder
2-3
106
20
0.1
2
Margarine
3-4
138
20
16
10
Milk powder
2-4
90
15
5
6
Red wine
3-5
100
15
97.4
15-20
Sunflower oil
10-14
138
20
0.1
2
Sugar
4-5
138
15
11.9
10
Milk
2-3
120
15
88
6-8
Flour
8-10
130
10
12.5
4-5
Cement
8-12
138
15
0.8
4-5
Paper
2-4
106
20
6.4
10
You will find further examples taken from actual use in our application manual,