04 Temperature adjustment
The temperature is measured at three points and can be adjusted at these
temperature points.
Before you do this, allow the device to cool down for at least 3 hours after the last
heating phase.
For preparation/installation of temperature calibration set( MLB-A12) , see
function "06 Temperature calibration“
Begin temperature adjustment:
Select by using the arrow keys (
) „
04 Temp.cal“
Press
key
Temperature calibration starts.
Temperature calibration of first point will take 8 min
after which you will hear an acoustic signal.
Enter the temperatur value as indicated on the
external thermometer with the help of the navigation
buttons
.
Confirm with
PRINT key, the second heating phase
is starting.
Temperature calibration of secend point will take 8
min after which you will hear an acoustic signal
Enter the temperatur value as indicated on the
external thermometer with the help of the navigation
buttons
.
Confirm with
PRINT key, the third heating phase is
starting.
Setup
29.12.04
Setup
Setup
29.01.08
Setup
Temperature calibration
1 point 14:59
Setup
29.01.08
Setup
Temperature calibration
1 point 00:00
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
25
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
24
Setup
29.01.08
Setup
Temperature calibration
2 point 14:59
Setup
29.01.08
Setup
Temperature calibration
2 point 00:00
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
120
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
122
Setup
29.01.08
Setup
Temperature calibration
2 point 00:00
1 point 07:59
2 pointt 07:59
90
89
3 point 07:59