KERN & SOHN MLS 50-3C User Manual
Page 62

62
MLS_C-BA-e-1330
7. How to enter a target weight for a sample
The user can enter a target weight stating a tolerance range for the sample.
This entry facilitates the application of identical sample sizes for each drying
process.
Press the arrow keys (
), to select “
Sample“
Press the
-key for the current setting to appear
Press the
-key; the current menu item blinks
Press the arrow keys (
), to select among the
settings below
•
None
Function disabled
•
Optional
A message appears giving information
whether the sample weight is within the
entered tolerance limits. The drying process
will be started even if the sample weight is
beyond the defined tolerance limits.
•
Obligatory
A message appears giving information
whether the sample weight is within the
entered tolerance limits. The drying process
will only be started if the sample weight does
not exceed the defined tolerance limits.
Confirm settings by pressing the
PRINT key.
After selecting
optional or obligatory enter the target
weight and tolerance limits.
Press the arrow keys (
), to select the parameters
Press the
-button and the current setting will flash
Press the arrow keys (
), to carry out the desired
setting
Confirm entry by pressing the
PRINT-key and enter
additional parameters as follows.