2 setting drying parameters – KERN & SOHN MLS 50-3C User Manual
Page 48

48
MLS_C-BA-e-1330
9.1.2 Setting drying parameters
You can enter parameters for each drying process in the menu, as follows:
• Heating profile (temperature, time)
• Shutoff criterion
• Unit of measured result
• Output interval
Operation
Display
To enable the operator menu, press the
Menu-key.
The first menu item “
Profiles“ appears.
1. Heating profiles
This is used to select suitable heating programs as well as to enter drying
temperature and drying time.
Press button
Press the arrow keys (
), to select among the
heating programs below (see also chpt. 11.8)
•
Standard
•
Step-by-step
•
Soft
•
Fast
First select a heating profile, then enter the desired
parameters (temperature/drying time).
Press the
-button; t
he cursor (►) marks the
adjustable parameter
Press the
-button anew and the current setting will
flash
Press the arrow keys (
), to carry out the desired
setting
Confirm by pressing the
PRINT key
For further parameter settings proceed as above