KERN & SOHN MLS 50-3C User Manual
Page 26

26
MLS_C-BA-e-1330
03 Calibration test
This is used to determine the deviation from the last adjustment. This is only a
check, i.e. no values are changed.
Press the arrow keys (
), to select “
03 Cal. test“.
Press
button
Ensure that there are no objects on the weighing
plate.
Press
PRINT key, on the display appears the weight
value of the required adjusting weight.
Then place the required adjustment weight in the
centre of the sample dish, press the
PRINT -key.
The result of the calibration test is shown on the
display.
Take away adjustment weight. To return to the menu
or weighing mode, press the
ESC-key several times.
Return to weighing mode:
Press the
ESC-key repeatedly until the query
”
SAVE?“ appears. Confirm query by pressing the
PRINT-key or reject it by pressing the ESC-key.
Setup
29.12.04
Setup
Einstellungen
Calibration check
Cear the pan
[Enter]
Einstellungen
Calibration check
Weight measuring
Einstellungen
Calibration check
Load weigh
[Enter]
t 50.0g
Einstellungen
Calibration check
Weight measuring
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
Setup
29.12.04
Setup