MLS_C-BA-e-1330
25
Ensure that there are no objects on the weighing
plate. Press the
PRINT key.
In the display there appears the value of the
adjustment weight.
Place the required adjustment weight in the centre of
the sample dish, then press the
PRINT key. If the
adjustment was successful, this will be followed by
automatic reset to the menu.
Take away adjustment weight.
The
ESC key may be used to cancel the adjustment.
Errors during adjustment or the use of an incorrect
adjusting weight will result in an error message.
Repeat adjustment.
Return to weighing mode:
Press the
ESC-key repeatedly until the query
”
SAVE?“ appears.
Confirm query by pressing the
PRINT-key or reject it
by pressing the
ESC-key.
Setup
!
Saveing...
Setup
29.01.08
We recommend:
Carry out all adjustments as close as possible to the maximum load of the balance.
Info about test weights can be found on the Internet at:
- TEE 150-1 (2 pages)
- TCB 200-1 (2 pages)
- CM 60-2N (14 pages)
- TAB 20-3 (14 pages)
- TEB 200-1 (11 pages)
- EMB 100-3 (18 pages)
- EMB 500-1GN (16 pages)
- EMB 500-1BE (48 pages)
- DBS 60-3 (91 pages)
- DLB 160-3A (59 pages)
- DLT 100-3N (77 pages)
- MLB 50-3C (68 pages)
- ABS 80-4N (96 pages)
- ABT 120-4M (72 pages)
- AES 100-4C (119 pages)
- AES 100-4N (85 pages)
- AET 500-4 (189 pages)
- ALS 160-4A (78 pages)
- ALT 160-4B (94 pages)
- ECB 10K5 (16 pages)
- ECE 10K5 (12 pages)
- FCB 3K0.1 (36 pages)
- FCE 3K1N (15 pages)
- FFN 1K-4N (32 pages)
- FKB 6K0.02 (42 pages)
- FKB 8K0.1A (39 pages)
- FKT 6K0.1 (53 pages)
- FKT 60K1L (7 pages)
- FOB 1.5K0.5 (21 pages)
- GAB 6K0.05N (41 pages)
- GAB 6K-5P (39 pages)
- WTB 1K-4N (35 pages)
- FOB 500-1S (17 pages)
- FOB 0.5K-4NS (27 pages)
- IXS 6K-3M (109 pages)
- IXS 6K-3M Platform (9 pages)
- SFB 50K-3XL (53 pages)
- SFB 50K-3XL Addendum (6 pages)
- KFE-TM (58 pages)
- FXN 3K-3 (22 pages)
- IEX 6K-3MEU (13 pages)
- IEX 6K-3MEU Safety (11 pages)
- OEX (15 pages)
- OEX Safety (12 pages)