50
MLS_C-BA-e-1330
Define
Settings are available for selection from 0 to 10
mg in 0.1 mg steps and from 5 seconds to 175 in 5
second steps.
Press
key; the cursor (►) marks the adjustable
parameter
Press the
- button anew and the current setting will
flash
Press the arrow keys (
), to carry out the desired
setting
Confirm by pressing the
PRINT key
Define 2
Drying is shut off as soon as the weight loss within
60 s is smaller than the set percentage value
(adjustable from 0.001% to 9.999%).
Press
key; the cursor (►) marks the adjustable
parameter
Press the
- button anew and the current setting will
flash
Press the arrow keys (
), to carry out the desired
setting
Confirm by pressing the
PRINT key
Test
Shut off for change of weighing value < 1mg in
freely defined time interval
Manual
Shutoff drying process manual
Temporal
Drying is shut off when set time has passed,
adjustable from 1min – 9h 59 min
Press
-
key; the cursor (►) marks the adjustable
parameter
Press the
- button anew and the current setting will
flash
Press the arrow keys (
), to carry out the desired
setting
Confirm by pressing the
PRINT key
%
60s
% 0.100%