1 daily inspection, Warning – Canon CXDI-55G User Manual
Page 23
5. Inspection and Maintenance
– 9 –
5.1.1 Daily Inspection
5.1.1.1 Before Turning ON the Power
5.1.1.2 After Turning ON the Power
WARNING
For safety reasons, be sure to turn OFF the power of each instru-
ment when the following inspections are going to be performed.
Otherwise, it may result in electric shock.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Ca
bl
e
Check that cables are not
damaged or cover of cables
is not torn.
Good/Bad
Good/Bad
Good/Bad
Contact Canon or distribu-
tor if there is any problem.
Check that the plugs and
locks of connectors are not
loose.
Good/Bad
Good/Bad
Good/Bad
Fully insert the cables and
lock them.
Check that there are no
cracks in the holder (pull-out
part) of the sensor cable.
Good/Bad
Good/Bad
Good/Bad
Contact Canon or distribu-
tor if there are any cracks in
the holder.
Se
ns
or
u
nit
Check that the cover or parts
are not damaged and not
loose.
Good/Bad
Good/Bad
Good/Bad
Contact Canon or distribu-
tor if there is any problem.
Inspection
Result
Remedy
Date
/
Date
/
Date
/
Gene
ra
l
Check that POWER lamp is
lit.
Good/Bad
Good/Bad
Good/Bad
Connect the power cable and
sensor cable properly.
Perform test exposure.
Good/Bad
Good/Bad
Good/Bad
Take measures if error mes-
sage is displayed by refer-
ring to the CXDI Series
Operation Manual. Contact
Canon or distributor if the
problem cannot be solved.
!