Optional linearisation of the scale curve – Flintec II IPC User Manual
Page 26

FlintWeigh II IPC Technical Manual, Rev. 1.06 November 2011
Page 26 of 36
8.6. Optional Linearisation of the Scale Curve
The standard scale calibration may be supplemented by a linearisation with up to 5 correction points. Hereby 
the correction points have to be in a stricly increasing order, which is: 
Min. capacity < Point 1 < Point 2 < Point 3< Point 4 < Point 5 < Max. capacity of the scale
Important: Before any linearisation you have to define the basic setup (see chapter 8.1), the scale parameters 
(see chapter 8.3 respective 8.4) and you have to calibrate the scale (see chapter 8.5). If you have to configure 
a multi scale system, then select the wished scale no. ("1" or "2") within the weight display (Fig. 7.1). 
 
1. Open the window “nWweigh Settings“ 
2. Open the register “Linearisation“ 
3. Correction point 1: A calibration weight will be 
put on the centre of the scale. The true applied 
weight as a multiple of the scale interval e has 
to be entered. Then you use the slider to correct 
the display value until it equals the calibration 
weight 
4. Define correction point 2, if applicable (this is
done like correction point 1 but uses a higher 
weight) 
5. Define further correction points, if applicable 
6. For saving the calibration data click the button 
“Save calibration data”. Now a warning appears 
which describes the consequences in legal for 
trade applications. (“Attention! After saving 
the calibration data, the internal calibration 
counter (TAC) is still counted forward. 
Afterwards the scale has to be re-approved 
by the Calibration Authority!”) 
7. For confirming the saving click the button
“Save”.
.
 
Hint: After the linearisation you have to close the window FlintWeigh II Settings by clicking the button “X“ (upper 
right window corner). In multi scale applications you may select and correct the second scale now. 
