Configurability, Signals and alarms, Technical da – EVCO EC6759 User Manual
Page 2

2
3
long as the instrument displays the label of the desired parameter.
To modify the parameter value keep pushed the key T3 (the instrument displays the actual
value) and at the same time push and release over and over the key T1 or T2 as long as the
instrument displays the desired value (keeping pushed the key T1 or T2 the value gets de-
creased or increased more quickly); after the modification release the key T3 last (to the
release of the key T3 the instrument displays the label of the parameter again).
To turn out from the configuration parameters setting procedure keep pushed at the same time
for four seconds at least the keys T1 and T2 or do not operate with the keys for fifty seconds
at least (time-out exit).
ADDITIONAL INFORMATIONS
-
for the whole period of a corrupted memory data alarm the access to the configura-
tion parameters setting procedure is refused
-
the modification of a parameter value which unit of measure is the hour or the
minute or the second has not immediate effect; to obtain this effect it must not be
executed during the course of the value
-
the modification of the parameters /6u, /7u, rP 1 ... 8 value has immediate effect; to
avoid damages to the connected users, it must be executed with the users not
connected
-
the modification of the parameter /Au value has not immediate effect; to obtain this
effect, after the modification turn OFF and turn ON again the instrument
-
the configuration parameters values get stored in a non volatile memory even if a
lack of power supply happens.
CONFIGURABILITY
CONFIGURABILITY
CONFIGURABILITY
CONFIGURABILITY
CONFIGURABILITY
CONFIGURATION PARAMETERS
LABEL MIN. MAX. U.M. ST.
PASSWORD
PA
-90
100 ---
0
password (§)
It is the password that permits to gain access to the second level.
LABEL MIN. MAX. U.M. ST.
MEASURE INPUTS
/1A
-9.9 +9.9 bar
0.0
probe 1 calibration (§)
It establishes a threshold to add algebraically to the signal coming from the probe 1 (for
instance to correct the signal).
/1b
-9.9 +9.9 bar
0.0
probe 2 calibration (§)
It establishes a threshold to add algebraically to the signal coming from the probe 2 (for
instance to correct the signal).
/2
0
3
---
0
digital filter
It establishes a time constant to apply to the signals coming from the measure inputs, as
indicated:
0 =
0 sec.
1 =
0.4 sec.
2 =
1.2 sec.
3 =
3.0 sec.
/5
0
2
---
1
decimal point
It establishes the decimal point position when a pressure gets displayed, as indicated:
0 =
the decimal point gets not displayed
1 =
the decimal point gets displayed on the dozens display
2 =
the decimal point gets displayed on the hundreds display.
/6A
-9.9 +99.9 points 0.0
lower end of scale for 4-20 mA input (it coincides with
the minimum calibration value of the transducer 1)
It establishes the lower end of scale for 4-20 mA input and it must coincide with the minimum
calibration value of the transducer 1.
/7A
-9.9 +99.9 points +7.0 upper end of scale for 4-20 mA input (it coincides with
the maximum calibration value of the transducer 1)
It establishes the upper end of scale for 4-20 mA input and it must coincide with the maximum
calibration value of the transducer 1.
/6b
-9.9 +99.9 points 0.0
lower end of scale for 4-20 mA input (it coincides with
the minimum calibration value of the transducer 2)
It establishes the lower end of scale for 4-20 mA input and it must coincide with the minimum
calibration value of the transducer 2.
/7b
-9.9 +99.9 points +7.0 upper end of scale for 4-20 mA input (it coincides with
the maximum calibration value of the transducer 2)
It establishes the upper end of scale for 4-20 mA input and it must coincide with the maximum
calibration value of the transducer 2.
/6u
-9.9 +99.9 bar
0.0
pressure value corresponding to 0 V of the analog output
It establishes the pressure value corresponding to 0 V of the analog output according with the
modality established with the parameter /Au.
/7u
-9.9 +99.9 bar
+99.9 pressure value corresponding to 10 V of the analog out-
put
It establishes the pressure value corresponding to 10 V of the analog output according with the
modality established with the parameter /Au.
/Au
0
1
---
0
pressure to associate to the analog output
It establishes the pressure to associate to the analog output, as indicated:
0 =
pressure read by the probe 1
1 =
pressure read by the probe 2.
LABEL MIN. MAX. U.M. ST.
ON-OFF PRESSURE REGULATOR ASSOCIATED RESPEC-
TIVELY TO THE FIRST WORKING SETPOINT AND TO THE
OUTPUT K 1, ... AND TO THE EIGHTH WORKING SETPOINT
AND TO THE OUTPUT K 8
r 0
+0.1 +9.9 bar
+5.0 hysteresis (differential) (§)
It establishes the hysteresis (differential) relative to the working setpoints.
rC1
-9.9 +99.9 bar
+10.0 first working setpoint
It establishes the pressure associated to the output K 1 according with the modality estab-
lished with the parameter rP1.
rC2
-9.9 +99.9 bar
+10.0 second working setpoint
It establishes the pressure associated to the output K 2 according with the modality estab-
lished with the parameter rP2.
rC3
-9.9 +99.9 bar
+10.0 third working setpoint
It establishes the pressure associated to the output K 3 according with the modality estab-
lished with the parameter rP3.
rC4 -9.9 +99.9 bar +10.0 fourth working setpoint
It establishes the pressure associated to the output K 4 according with the modality estab-
lished with the parameter rP4.
rC5
-9.9 +99.9 bar
+10.0 fifth working setpoint
It establishes the pressure associated to the output K 5 according with the modality estab-
lished with the parameter rP5.
rC6
-9.9 +99.9 bar
+10.0 sixth working setpoint
It establishes the pressure associated to the output K 6 according with the modality estab-
lished with the parameter rP6.
rC7
-9.9 +99.9 bar
+10.0 seventh working setpoint
It establishes the pressure associated to the output K 7 according with the modality estab-
lished with the parameter rP7.
rC8
-9.9 +99.9 bar
+10.0 eighth working setpoint
It establishes the pressure associated to the output K 8 according with the modality estab-
lished with the parameter rP8.
rd1
0
1
---
0
“high or low pressure” functioning of the output K 1
It establishes the output K 1 functioning, as indicated:
0 =
“high pressure” functioning
1 =
“low pressure” functioning.
rd2
0
1
---
0
“high or low pressure” functioning of the output K 2
It has the same significance of the parameter rd1.
rd3
0
1
---
0
“high or low pressure” functioning of the output K 3
It has the same significance of the parameter rd1.
rd4
0
1
---
0
“high or low pressure” functioning of the output K 4
It has the same significance of the parameter rd1.
rd5
0
1
---
0
“high or low pressure” functioning of the output K 5
It has the same significance of the parameter rd1.
rd6
0
1
---
0
“high or low pressure” functioning of the output K 6
It has the same significance of the parameter rd1.
rd7
0
1
---
0
“high or low pressure” functioning of the output K 7
It has the same significance of the parameter rd1.
rd8
0
1
---
0
“high or low pressure” functioning of the output K 8
It has the same significance of the parameter rd1.
rP1
0
1
---
0
pressure to associate to the output K 1
It establishes the pressure to associate to the output K 1, as indicated:
0 =
pressure read by the probe 1
1 =
pressure read by the probe 2.
rP2
0
1
---
0
pressure to associate to the output K 2
It has the same significance of the parameter rP1.
rP3
0
1
---
0
pressure to associate to the output K 3
It has the same significance of the parameter rP1.
rP4
0
1
---
0
pressure to associate to the output K 4
It has the same significance of the parameter rP1.
rP5
0
1
---
0
pressure to associate to the output K 5
It has the same significance of the parameter rP1.
rP6
0
1
---
0
pressure to associate to the output K 6
It has the same significance of the parameter rP1.
rP7
0
1
---
0
pressure to associate to the output K 7
It has the same significance of the parameter rP1.
rP8
0
1
---
0
pressure to associate to the output K 8
It has the same significance of the parameter rP1.
r 4
0
1
---
0
kind of hysteresis (kind of differential)
It establishes the kind of hysteresis (kind of differential), as indicated:
0 =
asymmetrical
1 =
symmetrical.
LABEL MIN. MAX. U.M. ST.
PRESSURE TO DISPLAY
/AL
0
1
---
0
pressure to display during the normal functioning
It establishes the pressure to display during the normal functioning, as indicated:
0 =
pressure read by the probe 1
1 =
pressure read by the probe 2.
LABEL MIN. MAX. U.M. ST.
RELAY OUTPUTS PROTECTION
C 0
0
999 sec. 0
disabling time to the outputs activation from the instru-
ment start
It establishes the time that disables the outputs activation from the moment of the instrument
start.
C 1
0
999 sec. 0
disabling time to the output activation from the previous
activation
It establishes the time that disables the output activation from the moment of the previous
output activation.
C 2
0
999 sec. 0
disabling time to the output activation from the previous
deactivation
It establishes the time that disables the output activation from the moment of the previous
output deactivation.
C 3 0
1
---
0
outputs status during a probe failure alarm
It establishes the status the outputs are forced during a probe failure alarm according
to the modality established with parameters belonging to family rP, as indicated:
0 = during a probe failure alarm the outputs will be forced OFF
1 = during a probe failure alarm output K8 will be forced ON; the remaining outputs
will be forced OFF.
ADDITIONAL INFORMATIONS
-
the symbol (§) indicates that the parameter is of the first level.
SIGNALS AND ALARMS
SIGNALS AND ALARMS
SIGNALS AND ALARMS
SIGNALS AND ALARMS
SIGNALS AND ALARMS
SIGNALS
If the LED L1 ... 8 is turned ON it means that the output K 1 ... 8 is activated.
If the LED L1 ... 8 flashes it means that a count of a disabling time to the output K 1 ... 8
activation is running (see the parameters C 0, C 1 and C 2).
If the instrument displays the indication “888” and the LED L1, ... and L8 are turned
ON it means that the autotest is running.
ALARMS
If the instrument displays the indication “E 2” (corrupted memory data alarm) it means that
there is a corruption of the configuration data in the memory (turn OFF and turn ON again the
instrument: if to the turning ON again the alarm does not disappear the instrument must be
replaced); during this alarm the access to the configuration parameters setting procedure is
refused, the outputs K 1 ... 8 get forced to the status OFF and the analog output signal is 0 V.
If the instrument displays the indication “E0A” (probe 1 failure alarm) it means that: the kind
of connected probe 1 is not proper (verify the kind of connected probe 1), the probe 1 is faulty
(verify the probe 1 integrity), there is a mistake in the instrument-probe 1 connection (verify
the instrument-probe 1 connection integrity), the pressure read by the probe 1 is outside the
limits permitted by the probe 1 in use (verify that the pressure near the probe 1 be inside the
limits permitted by the probe 1); during this alarm if the parameters rP1 ... 8 have value 0 the
outputs K 1 ... 8 get forced to the status established with the parameter C 3 and if the param-
eter /Au has value 0 the analog output signal is 0 V.
If the instrument displays the indication “E0b” (probe 2 failure alarm) it means that there is
one of the faults saw in the previous case but referred to the probe 2; during this alarm if the
parameters rP1 ... 8 have value 1 the outputs K 1 ... 8 get forced to the status established with
the parameter C 3 and if the parameter /Au has value 1 the analog output signal is 0 V.
If the instrument displays the pressure read by the probe 2 and the indication “E0A” every four
seconds (probe 1 failure alarm) it means that there is one of the faults saw in the previous
case but referred to the probe 1 and that the pressure read by the probe 1 is associated to a
relay output; during this alarm if the parameters rP1 ... 8 have value 0 the outputs K 1 ... 8 get
forced to the status established with the parameter C 3 and if the parameter /Au has value 0 the
analog output signal is 0 V.
If the instrument displays the pressure read by the probe 1 and the indication “E0b” every four
seconds (probe 2 failure alarm) it means that there is one of the faults saw in the previous
case but referred to the probe 2 and that the pressure read by the probe 2 is associated to a
relay output; during this alarm if the parameters rP1 ... 8 have value 1 the outputs K 1 ... 8 get
forced to the status established with the parameter C 3 and if the parameter /Au has value 1 the
analog output signal is 0 V.
TECHNICAL DA
TECHNICAL DA
TECHNICAL DA
TECHNICAL DA
TECHNICAL DAT
TT
T
TA
A
A
A
A
TECHNICAL DATA
Case:
plastic grey (PP0), self-extinguishing.
Size:
106 x 90 x 58 mm (4.17 x 3.54 x 2.28 in., 6 DIN modules).
Installation:
on DIN EN 50022 standard rail installation according with
DIN 43880 norms.
Type of protection:
IP 40.
Connections:
extractable screw terminal blocks with pitch 5 mm (0.19
in., power supply, measure inputs, relay outputs and ana-
log output) for cables up to 2.5 mm² (0.38 in.²).
Ambient temperature:
from 0 to +60 °C (+32 to +140 °F, 10 ... 90 % of not con-
densing relative humidity).
Power supply:
230 Vac, 50/60 Hz, 3 VA.
Insulation class:
II.
Measure inputs:
2 for 2 wires 4-20 mA pressure transducers (Ri 56 Ohm);
at terminal 87 and 91 +12 V (+50 %, -15 %) are available
to supply the transducers.
Working range:
configurable.
Setting range:
from -9.9 to +99.9 bar.
Resolution:
configurable for 0.1 or 1 or 10 bar.
Display:
3-digit display 12.5 mm (0.49 in.) high red LED display
with automatic decimal point and minus sign, relay out-
puts status indicators.
Outputs:
eight NO contact 6 (2) A @ 250 Vac relays for regulation
loads management and one 0-10 V analog output
configurable to be associated to the pressure read by the
probe 1 or 2.