beautypg.com

How to order – EVCO EC6294 User Manual

Page 3

background image

d1

0

1

---

1

kind of defrost

It establishes the kind of defrost that the instrument must manage, as indicated:

0 =

to resistances (electrical)

1 =

to hot gas (reversal of cycle).

d2

-55

+99 (*)

+2

defrost stopping setpoint

It is significant exclusively if the parameter /A has value 1, it establishes the temperature to

which a defrost gets stopped and it is referred to the evaporator probe.
d3

1

99

(**) 15

defrost maximum length

It establishes the defrost maximum length.
Passed the defrost maximum length from the moment in which the instrument activates a

defrost cycle, if the temperature read by the evaporator probe is below the setpoint established

with the parameter d2 the defrost ends.
d4

0

1

---

0

defrost cycle at the instrument start

It establishes if the instrument automatically must present a request of a defrost cycle at the

moment of the instrument start, as indicated:

0 =

inactive

1 =

the instrument automatically presents a request of a defrost cycle at the moment of the

instrument start.

d5

0

99

(**) 0

delay to a defrost cycle activation from the instrument

start

It is significant exclusively if the parameter d4 has value 1 or if the parameter i0 has value 1, it

establishes the delay with which the instrument automatically presents a request of a defrost

cycle from the moment of the instrument start and from the moment of the multifunction

digital input activation.
A manual presentation of a request of a defrost cycle during the count of the delay established

with the parameter d5 clears the delay.
d6

0

1

---

1

locking of the displayed temperature from a defrost cy-

cle activation

It establishes if to lock the displayed temperature from the moment in which the instrument

activates a defrost cycle, as indicated:

0 =

inactive

1 =

during a defrost, a dripping and as long as the temperature read by the cabinet probe

reaches the working setpoint, the instrument displays the temperature read by the

cabinet probe in the moment in which the instrument activates a defrost cycle.

d7

0

15

(**) 2

dripping length

It establishes the dripping length.
d9

0

1

---

0

disabling times to the compressor output activation over-

ride

It is significant exclusively if the parameter d1 has value 1, it establishes if to clear the disa-

bling times to the compressor output activation in the moment in which the instrument presents

a request of a defrost cycle, as indicated:

0 =

inactive

1 =

in the moment in which the instrument presents a request of a defrost cycle the

disabling times to the compressor output activation get cleared.

dA

---

---

(*)

---

evaporator probe reading (§)

It permits to display the temperature read by the evaporator probe.
db

0

1

---

0

times base

It establishes the unit of measure of the parameters expressed in time intervals, except for the

parameters L3 and n4, as indicated:

0 =

the unit of measure of the parameter d0 is the hour and the unit of measure of the

parameters C0, C1, C2, C5, d3, d5, d7, A3, A5, A6, A7, F5, i4 and i5 is the minute

1 =

the unit of measure of the parameter d0 is the minute and the unit of measure of the

parameters C0, C1, C2, C5, d3, d5, d7, A3, A5, A6, A7, F5, i4 and i5 is the second.

LABEL MIN. MAX. U.M. ST.

ALARM REGULATOR

A0

+1

+15 (*)

+2

hysteresis (differential)

It establishes the hysteresis (differential) relative to the parameters A1 and A2.
A1

-55

0

(*)

0

lower temperature alarm set relative to the working

setpoint

It establishes a threshold to add algebraically to the working setpoint; the value so obtained

establishes the temperature below which the lower temperature alarm gets activated and it is

referred to the cabinet probe.
If the parameter A1 has value 0 the lower temperature alarm never gets activated.
A2

0

+99 (*)

0

upper temperature alarm set relative to the working

setpoint

It establishes a threshold to add algebraically to the working setpoint; the value so obtained

establishes the temperature above which the upper temperature alarm gets activated and it is

referred to the cabinet probe.
If the parameter A2 has value 0 the upper temperature alarm never gets activated.
A3

0

240 (**) 120 disabling time to the temperature alarm activation from

the instrument start

It establishes the time that disables the temperature alarm activation from the moment of the

instrument start.
A5

-1

120 (**) 30

disabling time to the buzzer activation for the

multifunction digital input alarm indication

It is significant exclusively if the parameter i0 has values different to 0, it establishes the time

that disables the buzzer activation for the multifunction digital input alarm indication from the

moment in which the multifunction digital input alarm happens.
If the parameter A5 has value -1 during a multifunction digital input alarm the buzzer gets not

activated.
A6

0

240 (**) 5

disabling time to the temperature alarm activation

It establishes the time that disables the temperature alarm activation from the moment in

which the temperature alarm happens.
A7

0

240 (**) 15

disabling time to the temperature alarm activation from

the defrost cycle end

It establishes the time that disables the temperature alarm activation from the moment of the

6

3

defrost cycle end.
LABEL MIN. MAX. U.M. ST.

EVAPORATOR FANS REGULATOR ASSOCIATED TO THE

EVAPORATOR FANS OUTPUT

F1

-55

+99 (*)

-1

output deactivation setpoint

It is significant exclusively if the parameter /A has value 1 and if the parameter F7 has value 3

or 4, according with the modality established with the parameter F6, it establishes the tem-

perature to which the output gets deactivated and it is referred to the evaporator probe.
F2

+1

+15 (*)

+2

hysteresis (differential)

It is significant exclusively if the parameter /A has value 1 and if the parameter F7 has value 3

or 4, it establishes the hysteresis (differential) relative to the parameter F1.
F4

0

2

---

0

output functioning during the defrost and the dripping

It establishes the output functioning during a defrost and a dripping, as indicated:

0 =

during a defrost and a dripping the output gets forced to the status OFF

1 =

during a defrost and a dripping the output gets forced to the status ON

2 =

the output functioning during a defrost and a dripping gets established with the

parameter F7.

F5

0

15

(**) 2

evaporator fans stoppage length

It establishes the evaporator fans stoppage length.
F6

0

1

---

0

kind of output deactivation setpoint

It is significant exclusively if the parameter /A has value 1 and if the parameter F7 has value 3

or 4, it establishes the kind of output deactivation setpoint that the instrument must manage,

as indicated:

0 =

absolute (the output deactivation setpoint establishes the temperature to which the

output gets deactivated)

1 =

relative to the temperature read by the cabinet probe (the value obtained subtracting

the absolute value of the output deactivation setpoint to the temperature read by the

cabinet probe establishes the temperature to which the output gets deactivated).

F7

0

4

---

4

output functioning during the normal functioning

It establishes the output functioning during the normal functioning, as indicated:

0 =

during the normal functioning the output gets forced to the status OFF

1 =

during the normal functioning the output gets forced to the status ON

2 =

during the normal functioning the output gets forced to the status ON and during the

status OFF of the compressor output the output gets forced to the status OFF

3 =

during the normal functioning the output is associated to the setpoint established with

the parameter F1, it remains continuously activated as long as the temperature read by

the evaporator probe reaches the setpoint established with the parameter F1 and when

it falls below the setpoint established with the parameter F1 of the hysteresis value

(differential) established with the parameter F2 the output gets reactivated

4 =

during the normal functioning the output is associated to the setpoint established with

the parameter F1, it remains continuously activated as long as the temperature read by

the evaporator probe reaches the setpoint established with the parameter F1, when it

falls below the setpoint established with the parameter F1 of the hysteresis value

(differential) established with the parameter F2 the output gets reactivated and during

the status OFF of the compressor output the output gets forced to the status OFF

LABEL MIN. MAX. U.M. ST.

DIGITAL INPUTS

i0

0

5

---

4

action given during the multifunction digital input acti-

vation

It establishes the action given during the multifunction digital input activation, as indicated:

0 =

inactive

1 =

if the conditions permit it (the temperature read by the evaporator probe must be below

the setpoint established with the parameter d2) activating the multifunction digital input

the instrument presents a request of a defrost cycle

2 =

during the multifunction digital input activation the cabinet light output gets forced to

the status ON

3 =

during the multifunction digital input activation the compressor output, except what

established with the disabling times to the compressor output activation and the

cabinet light output get forced to the status ON

4 =

during the multifunction digital input activation the compressor and evaporator fans

outputs get forced to the status OFF and the cabinet light output gets forced to the

status ON

5 =

during the multifunction digital input activation the evaporator fans output gets forced

to the status OFF and the cabinet light output gets forced to the status ON.

If the parameter i0 has value 2, 3, 4 or 5 pressure on the key T5 during a multifunction digital

input alarm does not modify the cabinet light output status but it is indicated from the turning

ON (if before the multifunction digital input alarm activation the cabinet light output was deac-

tivated) or from the flashing (if before the multifunction digital input alarm activation the cabi-

net light output was activated) of the LED L5.
If the parameter i0 has value 3 or 4 the multifunction digital input alarm activation that happens

during a defrost cycle interacts on the compressor output status from the moment of the

defrost cycle end.
i1

0

1

---

0

kind of contact of the multifunction digital input

It establishes the kind of contact of the multifunction digital input, as indicated:

0 =

NO contact

1 =

NC contact.

i2

0

2

---

0

action given during the locking digital input activation

(®)

It establishes the action given during the locking digital input activation, as indicated:

0 =

inactive

1 =

during the locking digital input activation the compressor output gets forced to the

status OFF

2 =

during the locking digital input activation the compressor output gets forced to the

status OFF, from the moment of the locking digital input deactivation the compressor

output activation gets disabled for the time established with the parameter i5, in the

moment in which the compressor output gets reactivated the instrument increases a

locking digital input alarms counter, when the locking digital input alarms counter reaches

the number established with the parameter i3 the compressor output gets definitively

forced to the status OFF (instrument locking); to obtain the instrument unlocking turn

OFF and turn ON again the instrument.

Display:

3-digit display 12.5 mm (0.49 in.) high red LED display

with automatic minus sign, output status indicators, in-

strument status indicator.

Outputs:

five relays of which one 16 (2) A @ 250 Vac relay for one

1½ HP @ 250 Vac compressor (NO contact) management,

three 6 (3) A @ 250 Vac relays for evaporator fans (NO

contact), defrost system management (change-over con-

tact) and for cabinet light control (NO contact) and one

6 (3) A @ 250 Vac relay associated to the instrument sta-

tus (change-over contact).

Kind of defrost managed:

to resistances (electrical) and to hot gas (reversal of cy-

cle), automatic, manual and remote.

Defrost management:

for interval, stopping temperature and maximum length.

HOW TO ORDER

HOW TO ORDER

HOW TO ORDER

HOW TO ORDER

HOW TO ORDER

CODING SYSTEM
Instrument name:

EC 6-294.

Desired measure input:

P (for PTC/NTC probes).

Desired power supply:

220 (230 Vac).