Configurability, Signals and alarms, Technical data – EVCO EC4133 User Manual
Page 2: How to order
![background image](/manuals/675599/2/background.png)
pushed the key T1 or T2 the value gets decreased or increased more quickly): after the modi-
fication release the key T3 last (to the release of the key T3 the instrument displays the label PA
again); keep pushed at the same time for four seconds at least the keys T1 and T2 (passed four
seconds the instrument displays the first parameter of the second level).
To select a parameter of the second level push and release over and over the key T1 or T2 as
long as the instrument displays the label of the desired parameter.
To modify the parameter value keep pushed the key T3 (the instrument displays the actual
value) and at the same time push and release over and over the key T1 or T2 as long as the
instrument displays the desired value (keeping pushed the key T1 or T2 the value gets de-
creased or increased more quickly); after the modification release the key T3 last (to the
release of the key T3 the instrument displays the label of the parameter again).
To turn out from the configuration parameters setting procedure keep pushed at the same time
for four seconds at least the keys T1 and T2 or do not operate with the keys for fifty seconds
at least (time-out exit).
ADDITIONAL INFORMATIONS
-
for the whole period of a corrupted memory data alarm the access to the configura-
tion parameters setting procedure is refused
-
the modification of a parameter value which unit of measure is the hour or the
minute or the second has not immediate effect; to obtain this effect it must not be
executed during the course of the value
-
the configuration parameters values get stored in a non volatile memory even if a
lack of power supply happens.
CONFIGURABILITY
WORKING SETPOINT
LABEL MIN. MAX. U.M. ST.
WORKING SETPOINT
rA1
rA2
°C
0
working setpoint
It establishes the temperature associated to the output K 1.
CONFIGURATION PARAMETERS
LABEL MIN. MAX. U.M. ST.
PASSWORD
PA
-90
100
---
0
password (§)
It is the password that permits to gain access to the second level.
LABEL MIN. MAX. U.M. ST.
MEASURE INPUT
/0
01
31
---
(*)
kind of probe
It establishes the kind of probe that the instrument must recognize to its measure input, as
indicated:
01 = PTC probe
03 = NTC probe
10 = J thermocouple
11 = K thermocouple
20 = 3 wires Pt 100 probe 21 = 2 wires Pt 100 probe
30 = 2 and 3 wires 4-20 mA temperature transducer
31 = 2 and 3 wires 0-20 mA temperature transducer.
/1
-9
+10
°C
0
calibration (§)
It establishes a threshold to add algebraically to the signal coming from the measure input (for
instance to correct the signal).
/2
0
6
---
3
digital filter
It establishes a time constant to apply to the signal coming from the measure input, as indi-
cated:
0 =
0 sec.
1 =
0.4 sec.
2 =
1.2 sec.
3 =
3.0 sec.
4 =
8.0 sec.
5 =
19.8 sec.
6 =
48.0 sec.
/4
0
1
---
0
leading zeroes displaying
It establishes if to display the leading zeroes, as indicated:
0 =
the leading zeroes get not displayed
1 =
the leading zeroes get displayed.
/5
0
1
---
1
decimal point
It establishes the resolution with which the temperature gets displayed, as indicated:
0 =
the temperature gets displayed with the resolution of the unit of measure
1 =
the temperature gets displayed with the resolution of 1/10 of the unit of measure.
/6
-99
+999 points -20
lower end of scale for 0/4-20 mA input (it coincides with
the minimum calibration value of the transducer)
It establishes the lower end of scale for 0/4-20 mA input and it must coincide with the mini-
mum calibration value of the transducer.
/7
-99
+999 points +80
upper end of scale for 0/4-20 mA input (it coincides with
the maximum calibration value of the transducer)
It establishes the upper end of scale for 0/4-20 mA input and it must coincide with the maxi-
mum calibration value of the transducer.
LABEL MIN. MAX. U.M. ST.
ON-OFF TEMPERATURE REGULATOR ASSOCIATED TO THE
WORKING SETPOINT AND TO THE OUTPUT K 1
rA0
-99
+99
°C
(*)
hysteresis (differential) (§)
It establishes the hysteresis (differential) relative to the working setpoint.
rA1
-99
+999 °C
(*)
minimum working setpoint programmable
It establishes the minimum working setpoint programmable; the instrument automatically veri-
fies if the value established with the parameter rA1 is below the maximum working setpoint
programmable established with the parameter rA2.
rA2
-99
+999 °C
(*)
maximum working setpoint programmable
It establishes the maximum working setpoint programmable; the instrument automatically
verifies if the value established with the parameter rA2 is above the minimum working setpoint
programmable established with the parameter rA1.
rA3
0
1
---
1
cooling or heating functioning
It establishes the output functioning, as indicated:
0 =
cooling functioning
1 =
heating functioning.
rA4
0
1
---
0
kind of hysteresis (kind of differential)
It establishes the kind of hysteresis (kind of differential), as indicated:
0 =
asymmetrical
1 =
symmetrical.
rA5
0
1
---
0
locking of the working setpoint modification
It establishes if to prevent the modification of the working setpoint, as indicated:
0 =
the working setpoint can be modified
1 =
the working setpoint can not be modified as long as the parameter rA5 gets set to 0.
LABEL MIN. MAX. U.M. ST.
OUTPUT K 1 PROTECTION
CA0 0
999
sec. 0
disabling time to the output activation from the instru-
ment start
It establishes the time that disables the output activation from the moment of the instrument
start.
CA1 0
999
sec. 0
disabling time to the output activation from the previous
activation
It establishes the time that disables the output activation from the moment of the previous
output activation.
CA2 0
999
sec. 0
disabling time to the output activation from the previous
deactivation
It establishes the time that disables the output activation from the moment of the previous
output deactivation.
CA3 0
1
---
0
output status during a probe failure alarm
It establishes the status to which the output gets forced during a probe failure alarm, as indi-
cated:
0 =
during a probe failure alarm the output gets forced to the status OFF
1 =
during a probe failure alarm the output gets forced to the status ON.
CA4 0
1
---
0
disabling time to the output activation and deactivation
It establishes if to disable the output activation and deactivation for a fixed time, as indicated:
0 =
inactive
1 =
the output activation and deactivation get disabled for 3 sec.
LABEL MIN. MAX. U.M. ST.
TEMPERATURE ALARM REGULATOR ASSOCIATED TO THE
FIRST TEMPERATURE ALARM
AA0 (*)
+99
°C
(*)
hysteresis (differential)
It establishes the hysteresis (differential) relative to the parameter AA1.
AA1 -99
+999 °C
0
alarm setpoint
It establishes the temperature to which the temperature alarm gets activated according with
the modality established with the parameter AA4.
AA3 0
999
min. 0
disabling time to the alarm activation from the instru-
ment start
It establishes the time that disables the temperature alarm activation from the moment of the
instrument start.
AA4 1
7
---
1
kind of alarm
It establishes the kind of temperature alarm that the instrument must manage, as indicated:
1 =
the temperature alarm never gets activated
2 =
absolute lower temperature alarm
3 =
absolute upper temperature alarm
4 =
lower temperature alarm relative to the working setpoint
5 =
upper temperature alarm relative to the working setpoint
6 =
lower temperature alarm relative to the working setpoint with automatic
recomputation and reenabling
7 =
upper temperature alarm relative to the working setpoint with automatic
recomputation and reenabling.
LABEL MIN. MAX. U.M. ST.
TEMPERATURE ALARM REGULATOR ASSOCIATED TO THE
SECOND TEMPERATURE ALARM
Ab0 (*)
+99
°C
(*)
hysteresis (differential)
It has the same significance of the parameter AA0.
Ab1 -99
+999 °C
0
alarm setpoint
It has the same significance of the parameter AA1.
Ab3 0
999
min. 0
disabling time to the alarm activation from the instru-
ment start
It has the same significance of the parameter AA3.
Ab4 1
7
---
1
kind of alarm
It has the same significance of the parameter AA4.
LABEL MIN. MAX. U.M. ST.
CONNECTION IN A SERIAL NETWORK WITH EVCOBUS
PROTOCOL COMMUNICATION
L1
1
15
---
1
instrument address
It establishes the address to which the instrument (slave) answers when it is connected to a
serial network with EVCOBUS protocol communication managed from a master (for instance a
Personal Computer).
L2
0
7
---
0
instrument group
It establishes the group to which the instrument (slave) answers when it is connected to a
serial network with EVCOBUS protocol communication managed from a master (for instance a
Personal Computer).
ADDITIONAL INFORMATIONS
-
the symbol (§) indicates that the parameter is of the first level
-
the symbol (*) indicates that the value depends from the measure input presetting
requested
-
if the instrument is preset to accept at the measure input J/K thermocouples
the parameter /5 does not exist
-
if the instrument is not preset to accept at the measure input 2 and 3 wires 0/4-20
mA temperature transducers the parameters /6 and /7 do not exist.
SIGNALS AND ALARMS
SIGNALS
If the LED L1 is turned ON it means that the output K 1 is activated.
If the LED L1 flashes it means that a count of a disabling time to the output K 1 activation is
running (see the parameters CA0, CA1, CA2 and CA4).
If the instrument displays the indication
≡≡≡
≡≡≡
≡≡≡
≡≡≡
≡≡≡
it means that the working setpoint can not be
modified (see the parameters rA5).
ALARMS
If the instrument displays the indication E2 flashing (corrupted memory data alarm) it means
that there is a corruption of the configuration data in the memory (turn OFF and turn ON again
the instrument: if to the turning ON again the alarm does not disappear the instrument must be
replaced); during this alarm the access to the working setpoint setting and the configuration
parameters setting procedures is refused and the output K 1 gets forced to the status OFF.
If the instrument displays the indication E0 flashing (probe failure alarm) it means that: the
kind of connected probe is not proper (see the parameter /0), the probe is faulty (verify the
probe integrity), there is a mistake in the instrument-probe connection (verify the instrument-
probe connection integrity), the temperature read by the probe is outside the limits permitted
by the probe in use (verify that the temperature near the probe be inside the limits permitted by
the probe); during this alarm the output K 1 gets forced to the status established with the
parameter CA3.
If the instrument displays the indication E0C flashing (cold junction/third wire failure alarm)
it means that, if the instrument is preset to accept at the measure input J/K thermocouples
there is a mistake in the cold junction adjustment circuit (the instrument must be replaced), if
the instrument is preset to accept at the measure input 2/3 wires Pt 100 probes and it is set to
recognize to its measure input 3 wires Pt 100 probes the third wire of the probe is not con-
nected (verify the instrument-probe connection integrity); during this alarm the output K 1 gets
forced to the status established with the parameter CA3.
If the instrument displays the indication AL1 flashing alternated to the temperature read by
the probe (first temperature alarm) it means that the temperature read by the probe is outside
the limit established with the parameter AA1 (see the parameters AA0, AA1 and AA4); inactive.
If the instrument displays the indication AL2 flashing alternated to the temperature read by
the probe (second temperature alarm) it means that the temperature read by the probe is
outside the limit established with the parameter Ab1 (see the parameters Ab0, Ab1 and Ab4);
inactive.
If the instrument displays the indication 999 flashing (end of scale display) it means that the
temperature read by the probe is outside the working range permitted by the instrument (verify
that the temperature near the probe be inside the limits permitted by the instrument); inactive.
ADDITIONAL INFORMATIONS
-
the alarm codes are related in order of precedence.
TECHNICAL DATA
TECHNICAL DATA
Case:
plastic black (PPO), self-extinguishing.
Size:
48 x 48 x 104 mm (1.88 x 1.88 x 4.09 in.), with terminal
block.
Installation:
panel mounting, panel cutout 68 x 68 mm (2.67 x 2.67 in.),
with the equipped U-bracket.
Type of protection:
IP 54.
Connections:
extractable screw terminal block with pitch 5 mm (0.19
in., power supply and output) and screw terminal block
with pitch 5 mm (0.19 in., measure input) for cables up to
2.5 mm² (0.38 in.²).
Ambient temperature:
from 0 to +60 °C (+32 to +140 °F, 10 ... 90 % of not con-
densing relative humidity).
Power supply:
230 Vac or 115 Vac or 12-24 Vac/dc or 12 Vac/dc, 50/60
Hz, 2 VA.
Insulation class:
II.
Measure inputs:
1 configurable, hardware depending, for PTC/NTC probes
or J/K thermocouples or 2/3 wires Pt 100 probes or 2
and 3 wires 0/4-20 mA temperature transducers (Ri 56
Ohm); at terminal 15 +12 V (+30 %, -20 %) are available to
supply the transducer.
Working range:
from -50 to +150 °C (-58 to +302 °F) for PTC probe, from
-40 to +110 °C (-40 to +230 °F) for NTC probe, from -99 to
+700 °C (-99 to +999 °F) for J thermocouple, from
-99 to +999 °C (-99 to +999 °F) for K thermocouple, from
-99 to +600 °C (-99 to +999 °F) for 2/3 wires Pt 100 probes,
configurable for 2 and 3 wires 0/4-20 mA temperature trans-
ducers.
Setting range:
from -99 to +999 °C (-99 to +999 °F).
Resolution:
configurable for 0.1 (except the instruments preset to ac-
cept at the measure input J/K thermocouples) or 1 °C
(1 °F).
Display:
3-digit display 12.5 mm (0.49 in.) high red LED display
with automatic decimal point and minus sign, output sta-
tus indicator.
Outputs:
one change-over contact 8 (3) A @ 250 Vac relay for regu-
lation load management.
HOW TO ORDER
CODING SYSTEM
Instrument name:
EC 4-133.
Desired measure input:
P (for PTC/NTC probes),
J (for J/K thermocouples),
C (for 2/3 wires Pt 100 probes),
I (for 2 and 3 wires 0/4-20 mA temperature transducers).
Desired power supply:
220 (230 Vac)
115 (115 Vac)
024 (12-24 Vac/dc)
012 (12 Vac/dc).
Options:
custom configuration, temperature display in °F, green LED
display, SSR output, serial port.
2
3