EVCO EV9346J9 User Manual
Page 2
![background image](/manuals/675259/2/background.png)
EVCO S.p.A.
This document belongs to Evco; unless you are authorized by Evco, you can not publish it.
Via Mezzaterra 6, 32036 Sedico Belluno ITALY
Evco does not take any responsibility about features, technical data and possible mistakes related in this document or coming by its use.
Phone +39-0437-852468 • Fax +39-0437-83648
Evco does not take any responsibility about damages coming by the non-observance of the additional information.
[email protected] • www.evco.it
Evco reserves the right to make any change without prior notice and at any time without prejudice the basic safety and operating features.
PT • 35/09
If functioning with 2 measurement inputs:
top probe error
Remedies:
• the same as the previous case but relative to the top probe
Main consequences:
• the top output will be deactivated
• the acoustics output and the buzzer output will be acti-
vated
Pr2
floor probe error (only if functioning with 2 measurement
inputs)
Remedies:
• the same as the previous case but relative to the floor probe
Main consequences:
• the floor output will be deactivated
• the acoustics output and the buzzer output will be acti-
vated
When the causes of the alarm have disappeared, the instrument will go
back to normal functioning.
10
TECHNICAL DATA
10.1 Technical data
Container: grey self-extinguishing.
Front panel protection rating: IP 54.
Connections: removable terminal boards (power supply, inputs and
outputs), 6-pole connector (serial port).
Temperature of use: from 0 to 55 °C (from 32 to 131 °F,
10 ... 90% relative humidity without condensate).
Power supply: 115 ... 230 VAC, 50/60 Hz, 5 VA (approx) or 24 VAC.
Alarm buzzer: incorporated.
Measurement inputs: can be configured:
• 1 (chamber probe) for J/K thermocouple
if functioning with 1 measurement input
• 2 (top probe and floor probe) for J/K ther-
mocouple if functioning with 2 measure-
ment inputs
11
WORK SET-POINT, POWER DISTRIBUTED AND CONFIGURATION PARAMETERS
11.1 Work set-point
MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
r1
r2
°C/°F (1)
150
not visible
r1
r2
°C/°F (1)
not visible
150
r7
r8
°C/°F (1)
not visible
150
11.2 Power distributed
PARAM. MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
0
100
%
50
not visible
0
100
%
50
not visible
11.3 Configuration parameters
PARAM. MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
SP
r1
r2
°C/°F (1)
150
not visible
SP1
r1
r2
°C/°F (1)
not visible
150
SP2
r7
r8
°C/°F (1)
not visible
150
PARAM. MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
Po1
0
100
%
50
not visible
Po2
0
100
%
50
not visible
PARAM. MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
CA1
-25/-50 25/50 °C/°F (1)
0
0
CA2
-25/-50 25/50 °C/°F (1)
not visible
0
P0
0
1
- - - -
0
0
P2
0
1
- - - -
0
0
P5
0
(3)
- - - -
0
0
P6
0
(3)
- - - -
1
1
PARAM. MIN.
MAX.
U.M.
1 INPUT
2 INPUTS
r0
1
99
°C/°F (1)
5
5
r1
0
r2
°C/°F (1)
50
50
r2
r1
999
°C/°F (1)
350
350
r6
1
99
°C/°F (1)
not visible
5
r7
0
r8
°C/°F (1)
not visible
50
r8
r7
999
°C/°F (1)
not visible
350
PARAM. MIN.
MAX. U.M.
1 INPUT
2 INPUTS
c0
0
2
- - - -
0
not visible
c1
1
999
s
80
not visible
c2
0
3
- - - -
1
not visible
c3
0
99
°C/°F (1)
10
not visible
PARAM. MIN.
MAX. U.M.
1 INPUT
2 INPUTS
A1
0
999
°C/°F (1)
0
0
A2
0
240
min
0
0
A3
0
2
- - - -
0
0
A4
0
999
°C/°F (1)
not visible
0
A5
0
240
min
not visible
0
A6
0
2
- - - -
not visible
0
PARAM. MIN.
MAX. U.M.
1 INPUT
2 INPUTS
i1
0
1
- - - -
0
0
i5
0
2
- - - -
0
0
i6
0
1
- - - -
0
0
PARAM. MIN.
MAX. U.M.
1 INPUT
2 INPUTS
LA
1
247
- - - -
247
247
Lb
0
3
- - - -
2
2
LP
0
2
- - - -
2
2
(1)
the unit of measurement depends on parameter P2
(2)
set the parameters relative to the regulators appropriately after modification of parameter P2
(3)
the value depends on the type of functioning (1 with 1 measurement input and 3 with 2 measurement inputs)
(4)
the parameter differential is 10 °C/18 °F.
WORK SET-POINT
work set-point
top set-point
floor set-point
POWER DISTRIBUTED
power distributed to the top (percentage of c1); see also c0 and c1
power distributed to the floor (percentage of c1); see also c0 and c1
WORK SET-POINT
work set-point
top set-point
floor set-point
POWER DISTRIBUTED
power distributed to the top (percentage of c1); see also c0 and c1
power distributed to the floor (percentage of c1); see also c0 and c1
MEASUREMENT INPUTS
with 1 measurement input, chamber probe inset; with 2 measurement inputs,
top probe offset
floor probe offset
type of probe
0 = J
1 = K
temperature unit of measurement (2)
0 = °C
1 = °F
quantity shown by the upper part of the display during the on state or during
normal functioning
0 = with 1 measurement input, chamber temperature; with 2 measurement
inputs, top temperature
1 = with 1 measurement input, work set-point; with 2 measurement inputs, top
set-point
2 = temperature of the floor
3 = floor set-point
quantity shown by the lower part of the display during the on state or during
normal functioning
0 = with 1 measurement input, chamber temperature; with 2 measurement
inputs, top temperature
1 = with 1 measurement input, work set-point; with 2 measurement inputs, top
set-point
2 = temperature of the floor
3 = floor set-point
MAIN REGULATOR
with 1 measurement input, work set-point differential; with 2 measurement
inputs, top set-point differential
with 1 measurement input, minimum work set-point; with 2 measurement in-
puts, top minimum set-point
with 1 measurement input, maximum work set-point; with 2 measurement in-
puts, top maximum set-point
floor set-point differential
minimum floor set-point
maximum floor set-point
VARIOUS
restraint between the power distributed to the top and power distributed to the floor
0 = no restraint
1 = the modification of the power supplied to an output automatically causes
the supply of the maximum power to the other
2 = the modification of the power supplied to an output causes an automatic
adaptation of the power supplied to the other such to guarantee that the
sum of the two percentages is always 100
cycle time for the top output and floor output switch-on, see also Po1 and Po2
event that causes the activation of the rapid heating function
0 = function cannot be activated
1 = press
for 1s (make sure that the instrument is in on state and that no
procedure is in progress)
2 = pass from stand-by state to on state
3 = press
for 1s (make sure that the instrument is in on state and that no
procedure is in progress) or pass from stand-by state to on state
temperature of the chamber over which the rapid heating function is inter-
rupted (relative to the work set-point i.e. “work set-point - c3”)
TEMPERATURE ALARMS
with 1 measurement input, temperature of the chamber above which the cham-
ber temperature alarm is activated; with 2 measurement inputs, temperature of
the top above which the top temperature alarm is activated; see also A3 (4)
with 1 measurement input, chamber temperature alarm delay; with 2 measure-
ment inputs, top temperature alarm delay
with 1 measurement input, type of chamber temperature alarm delay; with
2 measurement inputs, type of top temperature alarm delay
0 = no alarm
1 = absolute (i.e. A1)
2 = with 1 measurement input, relative to the work set-point (i.e. “work set-point
+ A1”); with 2 measurement inputs, relative to the top set-point (i.e. “top set-
point + A1”)
floor temperature above which the floor temperature alarm is activated, se also
A6 (4)
floor temperature alarm delay
type of floor temperature alarm
0 = no alarm
1 = absolute (i.e. A4)
2 = relative to the floor set-point (i.e. “floor set-point + A4”)
DIGITAL INPUTS
on/stand-by input polarity
0 = live input active
1 = non-live input active
effect caused by the activation of the multifunction input
0 = no effect
1 = CHAMBER LIGHT SWITCH-ON/OFF - the activation of the input will cause the
chamber light to switch-on and the successive activation will cause its switch-
off
2 = BUZZER, ACOUSTIC OUTPUT AND BUZZER OUTPUT DEACTIVATION - the
activation of the input will cause deactivation of the buzzer, the acoustic
output and the buzzer output (activate the input again to deactivate these
utilities again)
type of contact of the multifunction input
0 = NO (input active with closed contact)
1 = NC (input active with open contact)
SERIAL NETWORK (MODBUS)
instrument address
baud rate
0 = 2.400 baud
1 = 4.800 baud
2 = 9.600 baud
3 = 19.200 baud
parity
0 = none (no parity)
1 = odd
2 = even
Digital inputs: 2 inputs:
• on/stand-by input in high voltage (230 VAC) with
configurable polarity
• multifunction input, for NO/NC contact (potential-
free contact, 5 V 1 mA).
Range of measurement: from -99 to 800 °C (from -99 to 999 °F) for
J thermocouple, from -99 to 999 °C (from -99 to 999 °F) for K thermo-
couple.
Resolution: 1 °C/1 °F.
Digital outputs: 6 relays:
• top (relay K1): 8 A res. @ 250 VAC (contact in
exchange)
• floor (relay K2): 8 A res. @ 250 VAC (contact in
exchange)
• alarm (relay K3): 8 A res. @ 250 VAC (NO
contact)
• chamber light (relay K4): 8 A res. @ 250 VAC
(NO contact)
• acoustics (relay K5): 8 A res. @ 250 VAC (NO
contact)
• on/stand-by (relay K6): 8 A res. @ 250 VAC
(NO contact)
The maximum current accepted on clamp 23 is 10 A.
Other outputs: buzzer output (12 V, max. 20 mA); the output is acti-
vated during alarms and errors, with continuous contribution.
Serial port: port for the communication with the supervising system
(through a serial interface, via TTL, with MODBUS communication
protocol) or with the programming key.