INT9 A/D converter
1 INT9 acquisition card
1 INT9 acquisition card
This manual describes the use of the INT9 A/D converter with the Clarity
software ver. 2.7 and later.
INT9 is an internal 24-bit A/D converter card that measures the voltages of
chromatographic and electrochemical detectors.
Fig 1: INT9
Brief technical overview
l
Up to 4 dataacquisition channels on one card (one and two channel
versions are also available).
l
The channels are completely independent sigma delta integrating A/D
converters, used for processing positive and negative voltages.
l
4 digital inputs.
l
8 digital outputs. The first four are also designed as relay contacts.
l
PCI Slot.
l
Excellent reliability and temperature stability.
The INT9 card of integration converters is a modernized successor of the
PCI card INT7.
1.1 A/D converter types in general
Prior to collecting and processing the chromatographic signal by a
computer, the analog signal of the detector must first be converted to
digital form using an analog/digital (A/D) converter.
In principle, there are three types of A/D converters – sampling,
integrating and δΣ (Delta-Sigma) - with continuous integration). The INT9
card uses Delta-Sigma integration.