Clarity INT9 User Manual
Page 22

4 Using the INT9 card
Clarity Hardware
filter) or 7.5, 15, 30, 60, 120 , 240 and 480 samples per second (for 60Hz
supply frequency filter).
Higher sampling rate allows for the measurement of narrower peaks, but it
also means a larger amount of data which will affect the size of the
resulting chromatogram and the speed of its processing.
Sample rate sufficient for successful peak detection is about 20 samples
per narrowest peak.
A higher sampling rate also increases the amount of noise and will need
setting the Peak Width parameter accordingly in the Integration Table on
the
Method Setup - Integration
tab.
Note:
Range and Sampling Rate are set individually for each detector.
- 18 -