Symptom
After completion of the 10 sec.
preparation time, the
displaycontinues to show
"overflow"

.


Possible Cause and Remedy

Check connection of
measuring cable. If
unsuccessful, the
resistance of the sample
may be too high (>19,99
M

).

Symptom
Fluctuating, unstable values are
displayed or considerable errors of
measurement occur.

Possible Cause and Remedy

Electric field effects can cause
this problem. A shielded low-
capacitance probe cable is
required.





















Repair Work
Please contact the BYK-Gardner
Service Department.