beautypg.com

Evaluation – ALTANA LC 2 User Manual

Page 11

background image

4.

Evaluation

The measured value constitutes
the resistance R [M

] of the liquid

between the electrodes (in the
annular gap). The cell constant
which depends on the design and
dimensions of the measuring cell
is 7.55 10

-3

cm

-1

.


The specific resistance, , and the
specific conductivity are
calculated from the following
equation:


= R • 132,5 cm

or [M • cm] =

R

[M

]


7,55

• 10

-3

cm

-1

[S • cm

-1

] =

7,55

• 10

-3

cm

-1

R

[M

]


with 1 = 1S

-1

= 1

1S





If the measurement result is
required for internal quality control
only, e.g. for entries in a check-
card, conversion is not necessary.