Start Up

The Main Menu is displayed when diagnostics are completed.

See section

5.1.3 Main Menu on page 16

for a detailed description.

4.4 Characteristics in continuous operation

After a continuous operating time of more than 24 hour without
switching off the instrument we recommend to perform a new
calibration with a distilled water cuvette/sample cell.

After a continuous operating time of more than 7 days without
switching off the instrument we recommend to switch off and on the
instrument to perform a new system check with filter adjustment
and lambda-calibration.