beautypg.com

Setting up, 2. installing the scale – A&D Weighing FS-30Ki User Manual

Page 7

background image

5

5. SETTING UP

5-1. Attaching a display pod to the display support column

1. Take out whole of the scale from the box

taking care not to pull the load cell cable.


2. Remove 2 clamps and 2 M4x12 screws.


3. Attach the display pod to the display

support column and tighten the
M4x12 screws removed at step 2.


4. Attach the clamps and tighten them

after tilting the display pod where
you like to use.

‰

Put the excess part of the load cell

cable into the display support
column

5-2. Installing the scale

1. Select the place for installing the scale.

Refer to “Cautions for installing the scale”
below.


2. Adjust the level of the base, using the spirit

level and leveling feet. There is an extra
foot under the display support column.
Adjust this foot to reach floor after adjusting
the level of the base.


3. Connect the main power cord to the outlet that

has the earth wiring. You may use the earth
terminal on the rear side of the display pod
to ground the scale.


4. If necessary, adjust the viewing angle of the

display by loosening the 2 clamps, changing
the angle and re-tightening the clamps.






Clamp

M4x12 screws

Clamp

Display pod

Spirit

level

Leveling foot

Clamps

Earth

terminal

Earth

terminal

Column

support foot

This manual is related to the following products: