A&D Weighing FS-30Ki User Manual
Page 25
23
7. To calibrate with the different weight, change the
displayed value using the 10-key pad.
If you hit the wrong number, press the C key. The
value returns to the capacity and enter again.
8. Place the calibration weight on the pan with the same
value as displayed, and wait until the STABLE
annunciator turns on.
9. Press the ENT key. The scale calibrate SPAN and
end will appear. Remove the weight from the pan,
and turns the power off.
Note
The value set in step 7 will be cleared after the power is switched off.
If the scale will suppose to move to another location, set the gravity
acceleration value for the current location and calibrate the scale
according to the procedure above. See the next section to set the
value.
10-2. Gravity acceleration correction
When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value of the scale to the value
of area where the scale will be used. Refer to the gravity acceleration map appended
to the end of this manual.
Note
It is not necessary to set the gravity acceleration correction when
calibrating the scale with a calibration weight at the place where it is
to be used.
1. At step 3. of the previous section “10-1. Calibration
using a weight”, enter a new value using 10-key pad.
The integral part “9” is fixed and enter the value after
the decimal point.
If you hit the wrong number, press the C key. The
value returns to the original number. Enter again.
2. Press the ENT key, then the Cal 0 will be
displayed.
3. If necessary to calibrate the scale using a calibration
weight, go to step 5 of “10-1. Calibration using a
weight”. To finish the setting, turn the power off.
-000
20.00
-00
end
00
-00
Cal
0
0
-00
9.8050