Resolution Switch
Action
To switch scale’s resolution, press
and
on remote controller in turn.
Function
The screen flashes the resolution to change in descending order, like 5kg, 2kg, 1kg
(available optional resolution ranges from 0.01 kg or lb to 50 kg or lb, depending on scale’s
capacity). After R
ESOLUTION
S
WITCH
, new resolution takes effect.
To balance between accuracy and measurement speed in high resolution mode,
A
NTI
‐M
OTION
L
EVEL
can be changed to fit your measurement application.
To learn how to set Anti‐Motion Level, please refer to the Anti‐Motion Level in 6
Advanced Operation section.
Condition
;
The scale must not be in H
OLD
mode. Otherwise, error message
will keep
flashing.
Battery Check
B
ATTERY
C
HECK
R
ESOLUTION
S
WITCH
NOTICE:
R
ESOLUTION
S
WITCHING
will change the apparent overloading range of
the scale. For example, if the resolution of a 3000kg scale is switched to
0.5kg, it will overload at a new weight of 3004.5kg (3000 + 9*0.5), while by
default, it overloads at a weight of 3009kg (3000 + 9*1).
NOTICE:
High resolution offers better accuracy at the cost of longer measuring
time and stricter requirement of load’s stability. Designed to meet the OIML
R76’s directive, this scale has the best (default) performance at 2000 to
3000 division.
NOTICE:
The default resolution will be restored next time when the scale is
powered on or enter System Setup Mode (no matter the setting is changed
or not). To save the changes in resolution for later, use S
YSTEM
S
ETUP
function, rather than R
ESOLUTION
S
WITCHING
.
NOTICE:
Setting or changing TARE has no effect on the ZERO setting.