3.7.1 Error Messages During Calibration

The following error messages might be displayed during the calibration of the tension
meter:

- The DISPLAY shows

EEE

.

The weight suspended from the process material is too heavy.

- The DISPLAY shows

-EEE

.

The weight suspended from the process material is too light.

You can restore the factory calibration any time with the following procedure:

Requirement:
-

Tension meter switched off.

To restore factory calibration:

-

Press and hold the MEM and POWER keys until the DISPLAY shows

E - 0

cN

.

- Release first the POWER key and then the MEM key.
-

Press and hold the DAMP and RECALL keys, then additionally press and hold the

MEM key until the DISPLAY shows

- - - -

cN

.

The factory calibration is restored.

-

Release the DAMP, RECALL and MEM keys

- Press the POWER key.

The instrument switches off.

3.7.2 Restoring the Factory Calibration