Quickstart Guide

The following instructions will demonstrate how to operate the Remote Display with one Digimatic

indicator and one quadrature Probe.

1.

Plug in the supplied power adapter to the Remote Display Mux Box (item 5 on page 3). The red

LED on the rear panel will light up.

2.

Power up the supplied Android tablet. Press and hold the On Button (approx. 5 seconds) until

HP logo appears on screen and then release the On button. The display will take approx. 30

seconds to startup. Using the supplied USB Micro to USB Type-A cable (item 1 page 3 on Back

Panel), connect the Android tablet to the Remote Display Mux Box. Do not connect the tablet to

the USB Type-B connector, as this will not function properly!

3.

The Remote Display application should automatically start on the Android tablet. If it does not

start, unplug the tablet from the gauge, wait a few seconds, and then reconnect.

4.

Plug a quadrature probe into channel 1

a.

The Android application defaults all channels to quadrature mode, so the gauge should

start reporting counts immediately after connecting. Confirm this by moving the spindle

of the probe.

5.

Plug the device (with Digimatic output) into channel 2

To configure channel 2 for Digimatic communication mode, long-press on the CH 02. Channel 2

Properties page will appear on tablet. S

oll do to the setti g fo Gage T pe, the ake su e

Digimatic is selected. Press the back button

on the tablet to return to the main display.

6.

Fo all Digi ati de i es, data a e a uall e uested p essi g the Re uest utto fo

the associated channel. Alternatively, data can be automatically updated at constant intervals

using the automatic update feature. To set an automatic update interval ( the Remote Display

will not update the Digimatic Data unless requested):

a.

Long-press on the appropriate channel (CH 01, CH 02, CH 03, or CH 04)

b.

Scroll down to the optio titled Auto- e uest i te al

c.

Select the desired update rate ( fastest rates may not be supported by all devices)

Note that the Re uest utto ill o lo ge e isi le fo that ha el. It is ot possi le to
manually request while the automatic function is active.