Parameter list – Checkline EMS4 User Manual

Page 13

background image

13

Parameter List

The following list summarizes the configuration parameters available for configuring each device

channel. When editing these values over RS-232 and/or USB, make sure to call the

Save command to

commit the values. Otherwise, they will be reset if the Remote Display is powered down.

Important Note: While these parameters can be changed over RS-232 and USB, the preferred way to

edit these settings is by using the

Channel Settings

menu on the Android application. While the Android

tablet is connected to the Remote Display, writing to these parameters will be blocked (reading is still

allowed). If parameter writing over the external data connection is required, the Android tablet must be

disconnected first.

Parameter

Description

Values

p003

Channel Enabled

0

– No

1

– Yes

p004

Units

1

– Inches

2

– Millimeters

3

– Hidden

p005

Measurement Mode

0

– Absolute (ABS)

1

– Total Indicated Runout (TIR)

2

– Incremental (INCR)

p006

Travel Direction

0

– Normal

1

– Reverse

p007

Hold Enabled

0

– Disabled

1

– Enabled

p008

Hold Mode

0

– Freeze

1

– Min

2

– Max

p009

Tolerance Enable

0

– Disabled

1

– Enabled

p910

Upper Tolerance

Enter value without decimal
Ex: 012345 = 0.12345

p911

Lower Tolerance

Enter value without decimal
Ex: 012345 = 0.12345

p914

Absolute Offset

Enter value without decimal
Ex: 012345 = 0.12345

p016

Gage Type

0

– Digimatic

1

– Encoder

p017

Glass Type
(encoders only)

1

– Inches

2

– Millimeters

p018

Base Multiplier
(one encoder tick)

5

– .

5

10

– .

100

– 0.001mm

200

– 0.002mm

p019

Ratio Mode Enabled

0

– Disabled

1

– Enabled

p020

Ratio Multiplier

Enter value without decimal