Parameter List
The following list summarizes the configuration parameters available for configuring each device
channel. When editing these values over RS-232 and/or USB, make sure to call the
Save command to
commit the values. Otherwise, they will be reset if the Remote Display is powered down.
Important Note: While these parameters can be changed over RS-232 and USB, the preferred way to
edit these settings is by using the
menu on the Android application. While the Android
tablet is connected to the Remote Display, writing to these parameters will be blocked (reading is still
allowed). If parameter writing over the external data connection is required, the Android tablet must be
disconnected first.
Parameter
Description
Values
p003
Channel Enabled
0
– No
1
– Yes
p004
Units
1
– Inches
2
– Millimeters
3
– Hidden
p005
Measurement Mode
0
– Absolute (ABS)
1
– Total Indicated Runout (TIR)
2
– Incremental (INCR)
p006
Travel Direction
0
– Normal
1
– Reverse
p007
Hold Enabled
0
– Disabled
1
– Enabled
p008
Hold Mode
0
– Freeze
1
– Min
2
– Max
p009
Tolerance Enable
0
– Disabled
1
– Enabled
p910
Upper Tolerance
Enter value without decimal
Ex: 012345 = 0.12345
p911
Lower Tolerance
Enter value without decimal
Ex: 012345 = 0.12345
p914
Absolute Offset
Enter value without decimal
Ex: 012345 = 0.12345
p016
Gage Type
0
– Digimatic
1
– Encoder
p017
Glass Type
(encoders only)
1
– Inches
2
– Millimeters
p018
Base Multiplier
(one encoder tick)
5
– .
5
10
– .
100
– 0.001mm
200
– 0.002mm
p019
Ratio Mode Enabled
0
– Disabled
1
– Enabled
p020
Ratio Multiplier
Enter value without decimal