3

1.Remove red protective rubber cap (if supplied) from probe.

2.Power-up Gage by pressing the center navigation button.

3.Place the probe FLAT on the surface to be measured. HOLD

STEADY. When a valid measurement is calculated, the Gage
BEEPS twice and the measurement is displayed.

4.Lift probe AT LEAST 2 INCHES (5cm) from the surface between

measurements - OR - leave probe on the surface in the same
location for continuous measurements.

Golden Rule

Golden Rule

Measure your uncoated part first! This quick zero-check
determines if a calibration adjustment is needed for your
substrate. (see pg.5)

Next, lay the included plastic shims onto a bare surface and
measure them individually to ensure the Gage measures a known
thickness within tolerance.

Operation Overview

Operation Overview

The PosiTector 6000 powers-up when the center navigation
button is pressed. To preserve battery life, the Gage powers-down
after approximately 5 minutes of no activity. All settings are
retained.

If memory is ON while continuous measurements are

being taken, only the last value on the display (when the probe is
lifted) is stored into memory. Scan (pg.17) stores ALL
measurements into memory.

NOTE:

Substrate

Current

Measurement

Battery Indicator

Typical Screen

Unit of

Measurement

Microns

F

56