Setup

Setup Menu

Setup Menu

Reset

16

NOTES:

Hi Res = OFF
Memory = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Cal Lock = OFF
Display = None
N Lock = OFF

Fast Mode = OFF
Scan Mode = OFF
Backlight = Night
SYNC = ON
USB = ON
Bluetooth = OFF

A more thorough Reset can be performed when the Gage is
powered down by simultaneously holding both the center and
(+) buttons until the Reset symbol appears. This is handy when
the Gage fails to power-up or operate properly. It performs the
same function as a menu Reset with addition of:

- Keep the Gage away from metal during a Reset.

- Date and Time are not affected by a Reset.

Units = microns
Flip Display = Normal
White on Black = OFF

Language = English
Battery Type = Alkaline

Reset restores factory settings and returns the Gage to a
known, out-of-the-box condition. It is handy when settings
have been changed, if the Gage behaves unusually, or if a
calibration adjustment is not possible.

The following occurs:

- all batches are closed and stored measurements, images and

batch names are erased.

- all calibration adjustments and Cal Memory are cleared and

returned to the Gage’s factory calibration settings.

- the factory calibration symbol appears on the display: It

disappears if a calibration adjustment is made by the user.

- menu settings are returned to the following: