I

N T RODU CT I ON

The CHECK•LINE

®

TI-25M Ultrasonic Thickness Gauge measures the wall

thickness of metals, glass, ceramics and many rigid plastics. This gauge uses
the “pulse-echo” principle of ultrasonic testing where a short ultrasonic sig-
nal is transmitted from the probe. The signal travels through the measure-
ment sample until it is reflected back towards the probe from the back side
of the material. The elapsed time for the complete cycle is measured and
converted to an accurate thickness reading.

The TI-25M can be used to measure the extent of corrosion on the opposite,
inaccessible side of the wall by using the “Subtractive Method.” When the
thickness of the original wall is known, subtracting the thickness reading
obtained from the TI-25M will determine the extent of corrosion at the point
of probe placement. If the original wall thickness is not known, test readings
should be made along a grid of equally-spaced points to obtain a profile of
the thickness readings. The smallest thickness reading will locate the area of
greatest concern.

The TI-25M is factory-set to measure steel. The gauge can be easily adjust-
ed for accurate thickness readings on other materials. Refer to sections 6.0
through 6.3. To change the measuring units from inch (factory setting) to
mm refer to section 4.4.

2 .0

P

RECAU T I ON S

Do not use the standard probe in applications where material temperatures
exceed 200 °F (100 °C) as the probe will be damaged. Special High
Temperature
Probes should be used. Consult factory.

Keep the gauge free of dust (especially metal powders, carbon, etc.) as they
will damage the gauge. Use a damp cloth to clean the gauge after use.
DO NOT USE CHECMICAL SOLVENTS OF ANY KIND.

– 23 –

1 0 .0 M

AT ERI AL

S

AFET Y

D

ATA

S

H EET

(M SDA)

Section 1— Product Identification

Product Name: TI-25M

Generic Name: Ultrasonic Couplant

Manufacturer: Electromatic Equpt. Co.

NFPA Hazardous Materials Identification System (est)
Health 0 Flammability 0 Reactivity 0

Section 2— Hazardous Ingredients

This material does not contain any ingredients having known health hazards in concentrations
greater than 1%. This material does not contain any known or suspected carcinogens.

Section 3 — Physical Data (nominal)

Boiling Point: >220°F

Freezing Point: <20°F

Vapor Pressure: N/A

Evaporation Rate: N/A

Specific Gravity: >1.0Z

Solubility in Water: complete

pH: 7.35 – 7.9

Acoustic Imp.: 1.726x10

6

Vapor Density: N/A

Appearance and Odor: water white, opaque gel; bland odor

Section 4 — Fire and Explosive Hazard Data

Flash Point: none Upper Exposure Limit: none Lower Exposure Limit: none
Special Fire Fighting Procedures: N/A

Extinguishing media: N/A

Unusual Fire and Explosion Hazards: none

Section 5 — Reactive Data

Stability: Stable Conditions to Avoid: none
Incompatibility (Materials to Avoid): none known
Hazardous Polymerization: will not occur
Hazardous Decomposition or Byproducts: none known

Section 6 —- Health Hazard and First Aid Data

Routes of Entry

1

:

Skin: not likely Ingestion: not normally

Eyes: not normally

Inhalation: no

Effects of Overexposure:

Acute: May cause temporary eye irritation

Chronic: none expected

First Aid Procedures:

Skin: Remove with water if desired.

Eyes: Flush with water for 15 minutes.

Ingestion: For large quantities, induce vomiting and call a physician Inhalation: N/A

Section 7 - Storage and Handling Information

Precautions to be taken in handling and storage: Store between 20 °F and 120 °F. Spills are
slippery and should be cleaned up immediately. Steps to be taken in case material is released or
spilled: Pick up excess for disposal. Clean with water. Waste disposal method: Dispose of in
accordance with federal, state, and local regulations.

Section 8 — Control Measures

Respiratory Protection: not required

Ventilation: not required

Protective Gloves: on individuals demonstrating sensitivity to TI-25M
Eye Protection: as required by working conditions

Other Protective Equipment: not required