Specifications 9. troubleshooting, Specifications, Troubleshooting – Checkline FGS-220VC User Manual
Page 32
- 27 -
8. Specifications
Model
FGS-220VC
Capacity
100kg (1,000N, 220lb)
Travel Speed
10-400mm/min (0.40-15.75
”/min)
Stroke
400mm (15.75
”)
Display
Dot-matrix LCD Four digit with sign
Operating Mode
MANU, JOG, SING, CONT, PROG
Communication
USB
Input
Over load from force gauge (The stand will stop immediately, if the over
load is detected.)
Measurement Table
150 x 200mm (5.91 x 7.87
”)
Operating Temperature
32-113 degrees Fahrenheit(0-45 degrees Centigrade)
Power
120VAC/240VAC
Weight
63kg (139lb)
Dimensions
300 x 885 x 500mm (11.81 x 34.84 x 19.69
”)
Accessories
USB cable
Communication cable for force gauge(FGV series)
PC Software
Free software is provided.
Available Force Gauge
FGV-series Force gauge
9. Troubleshooting
The following are general checkpoints; please call your local Shimpo representative or contact Shimpo
Instruments directly for further assistance.
Even if the power supply is turned on,
LCD doesn't display.
Confirm the voltage of the input supply.
The force gauge/load cell mounting
plate does not move.
Check power connections and power source, ensure that test
stand power is on.
Check whether the load hangs too much.
Check manual limit switches and adjust accordingly.
Confirm whether the emergency switch is ON.
Check the mounted force gauge is over load.
Check to see if the full travel range has already been achieved.
Check to be sure you are in the correct mode of operation.
The stand will not accept a program.
Move force gauge/load cell mounting plate to Home position.
Check to see if you are in the correct mode of operation (PROG).
The stand is noisy during operating.
If there is no abnormal condition, the sound of the motor and
gears may still be high due to rapid movement.