Program operation mode, Prog mode – Checkline FGS-220VC User Manual
Page 22
![background image](/manuals/656245/22/background.png)
- 17 -
5.4. Program operation mode
5.4.1. PROG mode
This mode of operation is ideal if user wants to operate the test stand with complex programmed movements.
Even if the user does not know the length of the object, the accurate measurement is possible because the
function to detect the object to be measured is provided.
Contents
The programmed data consists of a condition setting, starting point, a method of detection of the object, five
measurement points and a method of return.
The test stand will start to move when START button is pressed.
The stand will continue to move until one of the following events occurs: the STOP button is pushed, one of the
mechanical limit switches is tripped, the emergency reset button is pushed or the overload of the force gauge
is detected. If the force limit of the condition setting is detected, the present operation is interrupted and it will
jump to next point operation.
Start point
Program start point and return point
Detect point
Object detected point according to the condition
Standard point
Standard point for measurement
Point 1 to 5
Programmed point
Upper limit
Lower limit
Start point
Detect point
Standard point
Point 1
Point 4
Point 5
Point 3
Point 2
Example