Getting incorrect distances during a test
Re-enter all the test points distances (Remember that the distance is dependent on
previous set points and set points are accumulative).
Check if a previous or incorrect force limit is set in memory.
RS-232 output not working correctly
Check that proper cables are selected and attached to the force gauge test stand and PC.
Cable
DESCRIPTION
FGV-FGS250P
Communication cable between the test stand and the force gauge.
(for FGV-X and FGV-XY Models) (provided with stand)
FGS-50PCABLE
DFS cable for transferring data directly from the Shimpo DFS force gauge to PC.
FGV-RS232
Interface cable for data output from test stand to PC. (for FGV-X and FGV-XY Models)
FGS-ANALOG
Analog output cable for data acquisition requirements. (provided with stand)
FGS-100PVL
FGS-100PVH
Model
Low speed
High speed
Capacity
110 LB (500N, 50kgf)
5 - 120mm/min
20 - 500mm/min
Travel speed
2 types of speed switch
Stroke
140mm(with mechanical backlash 0.2 - 0.5mm)
Speed
5.0 -120.0mm/min
20.0 - 500.0mm/min
Distance
0.0 -140.0mm
Display
Repetition
number
0-9999
Operation mode
Manual, Inching, Single cycle, Continuous cycling, Program
mode,
Analog voltage
Output the displacement of the stand 10mV/mm MAX: ±1.4V
Output
signal
RS-232C
Yes
Input
Overload input
Stop the operation of the stand due to the overload of DFS,
FGV-X and FGV-XY
Alarm
Stop the motor due to the motor overload of the stand.
Height adjustment range
61-181mm (5 levels per 30mm)
Measuring table
150 × 145mm
Operation Temperature
0-45C (non condensing)
Power supply
AC115V
Weight
40 lbs (23 kg)
Standard Parts
120Vac outlet for force gauge, FGV-FGS250PV cable for
FGV-X and FGV-XY, FGS-Analog cable, gauge, mounting
screws, Allen wrench, wire guide, 6mm x 20mm all thread,
Optional
Grips, fixtures, data cables, digital and mechanical force gauges,
data collection software
8.0 Test Stand Specifi cations