Auto emission stop time setting (Function mode 3)

4.6.4

This setting is used to stop the emission automatically when no operation is performed for a certain
period of time.
The available setting time is within the range between 0 min and 120 min, which can be changed
by 1 minute.
When it is set to 0, the auto emission stop function is turned OFF and emission is performed
continuously.
When emission is stopped by the auto emission stop function, use the key or dial operations to
restart emission.

Operation

Display

Change the setting

value

Indicates alternately

* Previous setting value

CCW

CW

Go to function mode

4

Go to function mode 4

Display when emission is stopped by the auto emission stop function

Restart emission

Key or

dial operations