Emission duration mode
4.4.2
The emission time (ratio) can be set by 0.1º within the range between 0.1º/ 360º and 3.6º/ 360º in
the emission duration mode (DUR). (The emission time setting value shows the strobe emitting
time by angle while a rotating object rotates one revolution (360º).)
While the “UNIT” key is being pressed, the value converted from the emission angle to time can be
checked.
A longer emission time increases the brightness, but a measured object appears to be moving.
Also, a shorter emission time decreases the brightness, but a measured object appears to stand
more still.
Adjust the appropriate emission time according to your application and conditions and use the unit.
Press the MODE key to switch to the phase mode (PHA).
* When the external synchronous signal has not been entered, [ _ _ _ _ _ _ ] and the emission time are
indicated alternately.
When the external synchronous signal has been entered, only the emission time is indicated.
Operation
Display
Increase/decrease the
emission angle
Indicates the time
conversion value
while the key is being
pressed
Emission angle: Decreases
Emission angle: Increases
Indicates the value converted
from the emission angle to time
Indicates the value converted
from the emission angle to time
CCW
CW
Press the key to
switch to the phase
mode
Emission duration mode
Phase mode