Cannon Instrument D155 Density Meter User Manual
Page 39

36
5) Fit in the new pump tube around the rotor as
shown by pulling it out a little bit.
Fix the tube joint securely to the groove as
illustrated.
Rotor
Tube joint
Pump tube
Pull out
6) Match the axle and rotor, and fix the
cassette by turning clockwise until it stops.
Axle
Match
Pump tube
Fix the cassette by
turning clockwise
until it stops.
Caution!
Some samples may shorten the life of the pump tubing. Replace the
pump tubing when appropriate.
CANNON
®
D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA