background image

30

4-11. Parameters and default

Menu/Parameter

Description

Setting range

Default

0: Calibrate Cell

Calibration with pure water

1: View Stored
Results

Display measurement results

2: Set Sampling
Time

Select sampling time

00 to 99s

10s

3: Select Oil
Table

Select temperature compensation table for oil
23A: Crude Oils 60/60°F

23B: Refined Products 60/60°F
23D: Lube Oils 60/60°F
53A : Crude Oils 15°C

53B : Refined Products 15°C
53D : Lube Oils 15°C

59A: Crude Oils 20°C
59B: Refined Products 20°C
59D: Lube Oils 20°C

23A/23B/23D/
53A/53B/53D/

59A/59B/59D

53A

4: Set Test
Temperature

Set measurement temperature

15.0 to 25.0

°C

20.0

°C

5: Set Date &
Time

Enter date and time. Date format: MM/DD/YY
Time format: HH:MM (24 hour format)

01/01/00 00:00 to

12/31/99 23:59

6: Set LCD
Contrast

Adjust LCD brightness

0/25/50/75/100

50

7: Reset Test
Number

Reset Sample Test No. to ‘01’

01

8: Clear Stored
Results

Erase the memory of stored results. The D155 only
stores the last 10 measurement results.

9: Export Stored
Data to USB
Drive

Export Stored Data from Memory to an MS-Excel
formatted file.

10: Serial No. &
Firmware

View D155 Serial Number and Firmware Version


CANNON

®

D155 Density Meter Instruction & Operation Manual

Version 1.1b – August 2014; CANNON® Instrument Company