4. menu 4: set test temperature, 5. menu 5: set date and time – Cannon Instrument D155 Density Meter User Manual
Page 29

26
53D Lube Oils 15°C
59A
Crude Oils 20°C
59B
Refined Products 20°C
59D Lube Oils 20°C
4) Press [ENTER] to accept the selected table.
4-4. Menu 4: Set Test Temperature
Measurement temperature can be set as follows:
1) Press [MENU] then press [] or [] to select Menu 4 as shown below
2) Press [ENTER]
3) Press [] or [] key to select the digit to change
4) Press [] or [] to increase or decrease the number. The D155 allows any temperature between
15.0 and 25.0 °C
5) Press [ENTER] to accept the value entered
4-5. Menu 5: Set Date and Time
Enter the present date and time so that they are stored/printed with each measurement result.
1) Press [MENU] then press [] or [] to select Menu 5 as shown below
2) Press [ENTER]