beautypg.com

13 vacuum solution (optional), 4 .13 vacuum solution (optional) – BUCHI Multivapo P-12 User Manual

Page 36

background image

4 Description of function

36

Multivapor™ Operation Manual, Version E

4 .12 High-boiling solvents - Woulff bottle (optional)

To prevent boiling retardation and for high-
boiling solvents which tend to condense in the
vacuum tube as well as for solvents which tend
to foam an optional solvent reservoir – the so-
called Woulff bottle – can be fixed at the rear of
the instrument. The vapor is then first transferred
from the cover to the bottle and then further to
the condenser assembly.

Fig. 4.17: Woulff bottle installed

4 .13 Vacuum solution (optional)

Evaporation under vacuum is performed by
means of a vacuum pump. With the V-700 PTFE
diaphragm pump an ultimate vacuum of less
than 10 mbar is achieved, which is more than
sufficient for most applications.
Sophisticated vacuum control is gained using
the V-850 or V-855 controller. The latter includes
gradient functions, solvent libraries and auto-
matic vacuum control algorithms.

Fig. 4.18: Recommended vacuum solution for use with the Multivapor