8 troubleshooting, 1 malfunctions and their remedy, Malfunctions and their remedy – BUCHI Rotavapo R-215 User Manual
Page 51: 8troubleshooting, 8 .1 malfunctions and their remedy
![background image](/manuals/656047/51/background.png)
8 Troubleshooting
51
R-210/215 Operation Manual, Version F
8
Troubleshooting
This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a BUCHI technical
engineer who has access to the official service manuals. In this case, please refer to your local BUCHI
customer service agent.
8 .1
Malfunctions and their remedy
Table 8-1: General malfunction and their remedy
Malfunction
Possible cause
Corrective measure
Instrument does not work
Mains switch off
Switch on mains switch
Instrument is not connected to power
supply
Check if mains connection is okay
Fuse defective
Replace the fuse. If the malfunction
occurs again, contact the BUCHI
customer service.
Bath does not heat
Mains switch off
Switch on mains switch
Instrument is not connected to power
supply
Check mains connection
Over temperature protection was
activated
Let the bath cool down and empty it.
Set back the temperature probe by
pressing the button at the bottom
of the heating bath by means of a
flattended metal rod (e.g. a Torx or
hexagon wrench), Ø ~4 mm. See
Fig. 8.1
Fuse defective
Replace the fuse. If the malfunction
occurs again, contact the BUCHI
customer service.
Servo lift does not work
Various causes
Contact the BUCHI customer service
Flask does not rotate
Adjusting knob for rotation speed
at 0
Turn adjusting knob for rotation
speed clockwise until rotation starts
Restart instrument
Turn adjusting knob for rotation
speed to 0 then slowly turn clock-
wise until rotation starts
System is leaking
Joints have not been greased
Grease joints
Tube clips have not been fixed
correctly or are defective
Check tube clips
Tubes are leaky (brittle)
Replace tubes