5 Putting into operation
38
R-210/215 Operation Manual, Version F
5 .8 .2
Cooling water tubes
When connecting the white cooling water tubes,
consider the following:
• Use tube clips GL-14.
• The tubes used must all have the same inner
diameter (approximately 6 mm).
• For safety reasons, secure the tubes with
commercial tube pivoting clamps or cable
binders.
• To save cooling water, you can use the Distil-
lation Chiller F-1xx.
• Check the tubes from time to time and
replace them when they are brittle.
Fig. 5.9: Cooling water tubes
5 .8 .3
Vacuum tubes
When connecting the vacuum tubes, proceed
as follows:
• Use tube clips GL-14.
• The tubes used must all have the same inner
diameter (approximately 5 mm).
• Keep vacuum tubes as short as possible.
• When you are operating with the new
BUCHI Vacuum Controller V-850/855 and
Vacuum Pump V-700/710 connect a Woulff
bottle between the vacuum source and the
Rotavapor.
• When you are operating with a pump other
than a V-700/710, connect a valve unit to the
V-850/855 to control the vacuum.
• Tubes do not need to be secured.
• Check the tubes from time to time and
replace them when they are brittle.
Fig. 5.10: Vacuum tubes