beautypg.com

3 rear connections of the rotavapor, 2 motorized quick-action jack, 3 rear connections of the rotavapor 4.2 – BUCHI Rotavapo R-215 User Manual

Page 27: Motorized quick-action jack, 4 .2 motorized quick-action jack

background image

4 Description of function

27

R-210/215 Operation Manual, Version F

4 .1 .3

Rear connections of the Rotavapor

1

2

3

4

a Power supply for Rotavapor
b Socket for control cable to the vacuum

controller (RS-485) or the Vacuum Pump
V-700/710 (to be used alternatively)

c Socket for vapor temperature sensor
d Power supply for heating bath

Fig. 4.3: Rear connections of the Rotavapor

4 .2

Motorized quick-action jack

1

With the motorized quick-action jack, the eva-
porating flask can be lowered into, and raised
from, the water bath. This operation requires no
physical effort. Motion is effected by an integral
motor. The button a which can be pushed
up or down for the corresponding direction is
located on the handle of the quick-action jack.
When the instrument is switched off manually or
if a power failure occurs, the flask is automati-
cally raised from the heating bath.

Fig. 4.4: Quick-action jack

This manual is related to the following products: