Output Connections

The DV-III has two analog voltage outputs available for strip chart recording and one RS232 output
for computer or serial printer.

The analog voltages are:

0 - 1 volt DC corresponding to 0 to 100 Torque %
0 - 4 volts DC corresponding to -100

°

C to +300

°

C at 10 millivolts per

°

C

The location of the connector is shown in Figure G2:

Figure G2

The 9 pin connections are shown in Figure G3:

Figure G3

The analog Output Cable (Part No. DVP-96Y) plugs into the 9 pin connector in Figure G3 above. The
cable end wires with “spade lugs” are color coded as follows:

Red Wire:

Temperature Output

Black Wire:

Temperature Ground

White Wire: % Torque Output
Green Wire: % Torque Ground

Connector

RS-232 Serial Printer/Computer

Analog Output(s)

APPROVED POWER SUPPLY

TUV RHEINLAND

LICENSE NO. R9051346

RECOGNIZED COMPONENT

UL FILE NO. D 122103

NOTE

ANALOG OUTPUTS

ON SERIAL I/0

CONNECTOR

100-240 VAC 50/60 HZ 220VAC

MODEL DV-III BASE UNIT

240 CUSHING ST.

STOUGHTON, MA USA 02072

Transmit Data (TxD)

Receive Data (RxD)

Temperature and
% Torque Ground

% Torque

No Connection to Pin 1

Temperature

External Control

Serial Ground

External Control

1

2

3

4

5

6

7

8

9

Analog Output

Analog Output

Not Used