for usage in a 600 ml beaker is to return the material from the beaker back into the bottle. When using
smaller volumes in accessories such as Small Sample Adapter, UL Adapter or Thermosel, the fluid is
normally discarded.

Calibration Procedure for LV(#1-4) and RV,HA,HB(#1-7) Brookfield spindles:

1) Place the viscosity standard fluid (in the proper container) into the water bath.

2) Lower the DV-III into measurement position (with guard leg if LV or RV series Rheometer is

used).

3) Attach the spindle to the Rheometer. If you are using a disk shaped spindle, avoid trapping air

bubbles beneath the disk by first immersing the spindle at an angle, and then connecting it to the
Rheometer.

4) The viscosity standard fluid, together with the spindle, should be immersed in the bath for a

minimum of 1 hour, stirring the fluid periodically, prior to taking measurements.

5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate thermometer.

6) If the fluid is at test temperature (

+

/- 0.1

°

C of the specified temperature, normally 25

°

C),

measure the viscosity and record the Rheometer reading.

Note:

The spindle must rotate at least five (5) times before readings are taken.

7) The viscosity reading should equal the cP value on the viscosity fluid standard to within the

combined accuracies of the Rheometer and the standard (as discussed in the section entitled,
Interpretation of Calibration Test Results).

Calibration Procedure for a Small Sample Adapter

When a Small Sample Adapter is used, the water jacket is connected to the water bath and the water

is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies

with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction
manual.)

2) Place the sample chamber into the water jacket.

3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the DV-III.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test

temperature.

5) Measure the viscosity and record the Rheometer reading.

Note:

The spindle must rotate at least five (5) times before a viscosity reading is taken.

Calibration Procedure for a Thermosel System

When a Thermosel System is used, the controller stabilizes the Thermo Container at the test
temperature. A two-step process is recommended for the Thermosel: