Brookfield DV-III Rheometer User Manual
Page 40
- 40 -
Table D1 (continued)
Table D2 lists the model codes and spring torque constants for each Rheometer model.
The full scale viscosity range for any DV-III model and spindle may be calculated using the equation:
Where:
TK
=
DV-III Torque Constant from Table D2
SMC
=
Spindle Multiplier Constant listed in Table D1
SC4-27
27
25
0.34
SC4-28
28
50
0.28
SC4-29
29
100
0.25
SC4-31
31
32
0.34
SC4-34
34
64
0.28
SC4-37
37
25
0.36
CP40
40
0.327
7.5
CP41
41
1.228
2
CP42
42
0.64
3.8
CP51
51
5.12
3.84
CP52
52
9.83
2
Spiral Adapter
70
105
0.677
Thermosel DIN spindle
81
3.7
1.29
SSA DIN spindle
for 13R or 13RP chamber
82
3.75
1.29
SSA DIN spindle
for 7R or 7RP chamber
83
12.09
1.29
ULA DIN spindle
85
1.22
1.29
ULA DIN spindle
86
3.65
1.29
ULA DIN spindle
87
12.13
1.29
SPINDLE
ENTRY CODE
SMC
SRC
LVDV-III
0.09373
LV
2.5xLVDV-III
0.2343
4L
5xLVDV-III
0.4686
5L
1/4
RVDV-III
0.25
1R
1/2
RVDV-III
0.5
2R
RVDV-III
1
RV
HADV-III
2
HA
2xHADV-III
4
3A
2.5xHADV-III
5
4A
HBDV-III
8
HB
2xHBDV-III
16
3B
2.5xHBDV-III
20
4B
5xHBDV-III
40
5B
Table D2
MODEL
TK
MODEL CODE
ON DV-III SCREEN
Full Scale Viscosity Range [cP] = TK SMC
10,000
RPM
*
*