Brookfield DV-III Rheometer User Manual
Page 42
- 42 -
2
ω
Rc
2
Rb
2
χ
2
[Rc
2
– Rb
2
]
Shear Rate (1/sec) =
Where:
Rc = Radius of the container (in centimeters)
Rb = Radius of the spindle (in centimeters)
χ
=
Radius at which the shear rate is to be calculated (normally the same value as
Rb; in centimeters)
ω
=
Angular velocity of the spindle (Rad/Sec)
ω
=
N
2
π
60
*
N = Spindle speed in RPM
The procedure for entering SMC and SRC values is:
1) From the main screen, press
SELECT SPDL
. Note that the spindle select screen (Figure 7) will
display the last entered spindle.
2) Enter “99” using the numeric entry keys.
3) Press
SELECT SPDL
and Figure D1 is displayed.
Figure D1
Note that the values for the last entered SMC and SRC coefficients will be displayed at this
time if you had previously entered values for these coefficients.
4) Enter the SMC value.
5) Press the
NEXT
key to accept the SMC entry. The cursor will move to the SRC field.
6) Enter the SRC value.
7) Press the
NEXT
key to accept the SRC value (the cursor will move back to the SMC field).
8) Press the
SELECT SPDL
key to accept both entries. This will return you to the main screen
and SPINDLE 99 will be displayed. Please note that an attempt to press the
SELECT SPDL
key with invalid SMC and SRC entries results in an error message being displayed.