IV. THE OPTIONS MENU

The

OPTION

key is used to display the Options Menu screen. The Options Menu provides choices

of the following options:

OPTIONS MENU
1 PRINT 4 C -> F
2 ALARMS 5 CGS -> SI
3 DATA 6 HSK ON

Figure 19

1 - PRINT

set modes, when a serial printer is attached

2 - ALARMS

setting-high, low, and motor On/Off limits

3 - DATA

review of captured data

4 - C TO F

choice of temperature display

5 - CGS TO SI

choice of CGS or SI units

6 - HKS (ON/OFF)

handshake serial printer communication parameter

IV.1 Print Options

The DV-III is equipped with an RS-232 compatible serial output, which may be used with a serial
printer. The output characteristics are described in Appendix G.

The DV-III is designed to print with an X-on/X-off protocol, whereby the rheometer will not send
test data to the printer unless and until the printer sends it a signal saying that it is ready to receive
data. If the DV-III cannot tell that the printer is ready or connected, it assumes that the data has been
received and printed. Unfortunately, there is no design standard for the X-on/X-off communication
between printer and the transmitting device (DV-III). If you are not using a Brookfield serial printer,
you may have to experiment with the HSK ON/OFF setting to see if you can use the X-on/X-off
protocol. If possible, we suggest setting HSK to ON. To check your system, we suggest the
following:

1) Setup the printer and connect to the DV-III.

2) Set the DV-III to PRTN mode.

Press the

OPTION

key.

Press

NUMBER 1

in the Options Menu.

Press

NUMBER

3 to set PRTN.

3) The screen will change to the Options Menu.

4) Press

NUMBER 6

to read “6 HSK On”.

5) Press the

CANCEL

key to return to the main screen.

6) Enter a motor speed.

7) Press the

NEXT

key.

This should cause the printer to print one line of data. If there is no communication between the DV-