Brookfield DV-III Rheometer User Manual
Page 16
- 16 -
Speed Set Usage Notes:
1) When a speed set is used, all the captured test data is stored in memory until another speed set
test is run. When the next (either a new or memory resident) speed set test begins, the last
captured test data is over-written.
2) After a speed set is executed, the data is retained in memory by the DV-III even if the power
switch is turned off.
3) To halt a running program, press the
PROG RUN
key. Spindle rotation will stop and the
following screen will be displayed:
Figure 15
Press
PROG RUN
to continue the program where you left off. The step time will pick up at the
time it halted. Press the
CANCEL
key to stop the program.
III.5
Operator Lock-Out
Operator Lock-Out is used in conjunction with a speed set program that is loaded and ready to run.
It locks out most of the operating functions, leaving the operator with access to the
SELECT DISP,
MOTOR ON/OFF
,
PROG RUN
,
NEXT
and
CANCEL
keys. This reduces the chance for errors in
data gathering. Only the speed set program, which you choose for lockout, will run. You cannot
inadvertently change the current spindle entry and thus have incorrect Viscosity, Shear Rate or Shear
Stress data.
The procedure for setting Lock-Out is as follows. Start from the main screen after power up,
Autozero and Select SPDL entry have been completed:
1) Press the
PROG SPEED
key.
2
Select a speed set.
3) Press the number 3 key to use the speed set (as shown in Figure 11).
4) Press
PROG RUN
.
5) Press
OPTION
and the Lock-Out Options screen will be displayed.
Figure 16
6) Press the
NUMBER 1
key to Enable Lock-Out and the screen changes to the step sequence
screen (Figure 14) ready for you to select your step sequence.
7) Press
PROG RUN
and the program begins running in Lock-Out mode. Note that the top line
of the DV-III main screen is changed to display the letters LK indicating that the DV-III is