Quick-Check Red User Instructions

7.2 EDX PSU

Problem

Possible Cause

Solution

Quick Check powers on
momentarily and turns off

Low battery

Bad keypad

Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.

Quick Check does not power
on

Low battery

Bad keypad

Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.

Batteries installed
backwards or no spring
contact

Insure that positive terminals of both batteries (nub) face inward –
towards the black cap. Check that spring is attached to the battery cap.

Software reset

Remove battery cap & reinstall after one minute. Attempt to turn power on
again.

Display contrast too light Hold the Right Arrow key down while pressing the F2 key several times to

increase the display contrast. If nothing occurs, release both keys. Press
the power button and try again.

EDX PSU Power Supply Remove Power Supply and install batteries to see if the Quick Check or

Communicator II will power up. Return EDX PSU to factory for repair/
troubleshooting.

Display is completely dark

Display contrast too dark Hold the Arrow key down while pressing the F1 key several times to

decrease the display contrast.

Quick Check does not
appear accurate

Check installation &
system

Insure that shackles are in good working condition and aligned straight.
Verify system is applying force directly through the dynamometer with no
off center or torsional loads being applied to the instrument.

Local gravitational
variances

If being compared against dead-weights, check your local gravitational
constant. Use custom units to compensate or calibrate on-site.

Check repeatability

Place Quick Check in low-resolution mode. Lift an arbitrary weight several
times as close to capacity as possible. Record each weight reading. Do
the readings differ from each other? Calculate the standard deviation of
the readings using a spreadsheet such as Microsoft Excel. See if the
deviation is greater than 0.1% of the instrument capacity.

Compare against a
reference load.

Place Quick Check in low-resolution mode. Apply a known load near
instrument capacity. Check calibration date.

Radio communications not
working at all

Dead batteries. Distance
is excessive, dead radio
pocket

Bring remote closer to dynamometer. Allow several seconds to retrain.

Operating channels mis-
matched

Remote and link must be on the same operating channel. See Quick
Check and Communicator configurations of COM1 for radio (under Comm
menu) and Common Configurations.

Excessive radio noise or
interference in
environment

Remove dynamometer and Communicator from the environment. Attempt
communications in an area free of local radio signals. See Radio
Information section of the manual.

Remote reading changes to
dashes

Low batteries, lost
communications

See steps above for improving communications.

Display locks up on
DILLON marquee

Poor connection
between Quick Check
and Communicator

Remove batteries from Quick Check and Communicator, replace them
and power up.