2 edx psu, Edx psu – Dillon Quick-Check Wire Tension Meter (Red) User Manual
Page 34
34
Quick-Check Red User Instructions
7.2 EDX PSU
Problem
Possible Cause
Solution
Quick Check powers on
momentarily and turns off
Low battery
Bad keypad
Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.
Quick Check does not power
on
Low battery
Bad keypad
Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.
Batteries installed
backwards or no spring
contact
Insure that positive terminals of both batteries (nub) face inward –
towards the black cap. Check that spring is attached to the battery cap.
Software reset
Remove battery cap & reinstall after one minute. Attempt to turn power on
again.
Display contrast too light Hold the Right Arrow key down while pressing the F2 key several times to
increase the display contrast. If nothing occurs, release both keys. Press
the power button and try again.
EDX PSU Power Supply Remove Power Supply and install batteries to see if the Quick Check or
Communicator II will power up. Return EDX PSU to factory for repair/
troubleshooting.
Display is completely dark
Display contrast too dark Hold the Arrow key down while pressing the F1 key several times to
decrease the display contrast.
Quick Check does not
appear accurate
Check installation &
system
Insure that shackles are in good working condition and aligned straight.
Verify system is applying force directly through the dynamometer with no
off center or torsional loads being applied to the instrument.
Local gravitational
variances
If being compared against dead-weights, check your local gravitational
constant. Use custom units to compensate or calibrate on-site.
Check repeatability
Place Quick Check in low-resolution mode. Lift an arbitrary weight several
times as close to capacity as possible. Record each weight reading. Do
the readings differ from each other? Calculate the standard deviation of
the readings using a spreadsheet such as Microsoft Excel. See if the
deviation is greater than 0.1% of the instrument capacity.
Compare against a
reference load.
Place Quick Check in low-resolution mode. Apply a known load near
instrument capacity. Check calibration date.
Radio communications not
working at all
Dead batteries. Distance
is excessive, dead radio
pocket
Bring remote closer to dynamometer. Allow several seconds to retrain.
Operating channels mis-
matched
Remote and link must be on the same operating channel. See Quick
Check and Communicator configurations of COM1 for radio (under Comm
menu) and Common Configurations.
Excessive radio noise or
interference in
environment
Remove dynamometer and Communicator from the environment. Attempt
communications in an area free of local radio signals. See Radio
Information section of the manual.
Remote reading changes to
dashes
Low batteries, lost
communications
See steps above for improving communications.
Display locks up on
DILLON marquee
Poor connection
between Quick Check
and Communicator
Remove batteries from Quick Check and Communicator, replace them
and power up.