Quick-Check Red User Instructions
19
Setup:
Setup allows the user to configure how the Quick Check stores data internally. This
stored data can be downloaded via the 4 pin Lemo to a PC via a keyboard wedge or
other device.
l
Log Mode:
Disable: Turns the log feature off.
On Print: Used when the operator wants to do a lift and store of the force
and peak.
Auto ID Increment: Enable/Disable
This will increment the ID by 1 every time the "PRINT" key is
pressed and store it with each force and peak.
Auto ID Prompt: Enable/Disable (will only appear if Auto ID
Increment is Disabled)
If "Enabled", the operator will be prompted to key in an ID before
storing that ID with the force and peak.
Comm Port:
This allows you to choose where to export the "ON PRINT" logged
data.
a. Cell = Setup Cell for data export (4 pin Lemo)
b. Com1 = Not used
c. RADIO = Future development
d. Disabled = to turn off
On Lift: Used when the operator wants the load to be automatically
stored, once the load becomes stable. Note: additional filtering may be
required, as the load must be steady before the load will be recorded.
Maximum Records = 255
Threshold %: (based on capacity)
This is the load above where the unit will record the stable load.
Re-arm: (based on capacity)
This is the load the force must drop below before the "ON LIFT"
will reset (re-arm) and store another load.
Auto ID Increment: Enable/Disable
This will increment the ID by 1 every time the "PRINT" key is
pressed and store it with each force and peak.
Auto Send: Enable/Disable
Can export the readings live if enabled.
Note: All Log Modes will record Date and Time along with it's force and peak reading.