30

Quick-Check Red User Instructions

5 Changing Sheaves

As you use the Quick-Check on different diameter cables you must change to the
correct sheave size. To change sheaves, remove the hex head screws pointed out in
Figure

5.1

below. Replace the sheaves with the correct letter sheave and reinsert the

screws and tighten.

Insure sheaves installed agree with sheaves noted in the Wire calibration.

Exception: Sheaves match the wire diameter of the cable to be measured and
alternate calibration is selected as per the section Calibration to Specific Wire Type

on

page 31

.

Insure that the wire rope is riding in the groove of all three sheaves.

Figure 5.1 Changing Sheaves

Do not use the Quick-Check with cable larger than indicated on the sheaves.
Overload and damage to the instrument may result.

Do not mix sheave sizes. This will result in inaccurate measurement and possible
overload.

Screws