background image

GTX-Plus Series User’s Manual REV 2.1

22

Figure 16 Date&Time Menu

Press UP and DOWN to change the value, press and hold to scroll

values. Press LEFT and RIGHT key to move the cursor, press and hold to
scroll move the cursor. Press ENTER key to accept the value and return

to main menu page.


5) DIAGNOSTIC This use to check status of the load cell. If you

suspect that your load cell transducer has sustained an overload it is

possible to check the status of the load cell immediately.

Place the gauge horizontally on the flat level surface and go to

main menu page._Use UP and DOWN key to move the cursor point to

DIAGNOSTIC and press ENTER key the display will show Diagnostic menu

page. Press ESC to return to main menu page.

Last date of calibration.

Total overload coun

% offset when last calibrate.

% current offset.

Figure 17 Diagnostic Menu

If the % offset is between 5% - 10 % please contact your

supplier to arrange a recalibration of your gauge.


If the % offset is greater than 10% please contact your supplier

to arrange for load cell replacement.

These values are given as an indicator only – the need for

calibration/repair may vary according to the individual characteristics of
the load cell.


SET DATE&TIME

mm/dd/yy : 09/19/05

hh:mm:ss : 15:32:26


DIAGNOSTIC

LAST CAL. : 12/09/05

OVERLOAD COUNT: 10

ORG. OFFSET : +0.4 %
CUR. OFFSET : +0.4 %