Dillon GTX Plus Force Gauge User Manual
Page 22
GTX-Plus Series User’s Manual REV 2.1
22
Figure 16 Date&Time Menu
Press UP and DOWN to change the value, press and hold to scroll
values. Press LEFT and RIGHT key to move the cursor, press and hold to
scroll move the cursor. Press ENTER key to accept the value and return
to main menu page.
5) DIAGNOSTIC This use to check status of the load cell. If you
suspect that your load cell transducer has sustained an overload it is
possible to check the status of the load cell immediately.
Place the gauge horizontally on the flat level surface and go to
main menu page._Use UP and DOWN key to move the cursor point to
DIAGNOSTIC and press ENTER key the display will show Diagnostic menu
page. Press ESC to return to main menu page.
Last date of calibration.
Total overload coun
% offset when last calibrate.
% current offset.
Figure 17 Diagnostic Menu
If the % offset is between 5% - 10 % please contact your
supplier to arrange a recalibration of your gauge.
If the % offset is greater than 10% please contact your supplier
to arrange for load cell replacement.
These values are given as an indicator only – the need for
calibration/repair may vary according to the individual characteristics of
the load cell.
SET DATE&TIME
mm/dd/yy : 09/19/05
hh:mm:ss : 15:32:26
DIAGNOSTIC
LAST CAL. : 12/09/05
OVERLOAD COUNT: 10
ORG. OFFSET : +0.4 %
CUR. OFFSET : +0.4 %