To set DROP LEVEL, Press UP and DOWN to move the cursor

point to DROP LEVEL and press ENTER key The display will show the Set

Drop level menu page. Press ESC key to return the setup menu page.

DROP LEVEL MENU

DROP = 2.5 N


Press Zero key to
Reset value.

Figure 9 Set Percent Drop Menu

Use UP and DOWN keys to change the value, press and hold to

scroll values. Use RIGHT ARROW key to change unit. When the correct

value is reached press ENTER to save setting and return to setup menu

page.

*1st Peak feature will automatically disabled if you set DROP = 0 N.

example GTX 100N has drop level =20N. If the peak load before
sample break is 50 the load must drop to 30N in order for the GTX to

detect a 1st peak of 50N. If load continues to be applied above 50N (e.g.

to 75N), the GTX will return 75N as PEAK and 50N as 1st peak.

GTX capacity

100 N


80 N

Ultimate Peak

60 N

First Peak

40 N

20N Drop

Trigger Threshold

20 N

Time

0

Figure 9a